Troubleshooting
Network Troubleshooting
Issue
possible cause
possible solution
no communication to
networked unit(s)
Are master/slave menus
properly configured?
See
Networking
- page 93.
Is termination relay en-
abled in 1st & last 5200A on
network?
See
Termination Relay
-
page 95.
Is proper communication
method configured and
wired properly?
See
Networking
- page 93.
See
Wire Communication
Equipmen
t - page 57.
Are slaves enabled and recog-
nized at master?
enable slaves at
Menus →
System → Communications
→ Network →Master
more than one device is
configured as a master
verify configuration at
Menus → System → Com-
munications → Network
→Master/Slave menus on
all nodes.
slow network perfor-
mance
nicked wires;
improper wiring
Verify network wiring -
See
Wire Network
- page 52
5200A Systems Troubleshooting
Issue
possible cause
possible solution
time and date lost at
power up
bad lithium battery
replace lithium battery
no events in event log
event logging not enabled
See
Event Logging
- page
104.
no data in data log
data log interval time has not
passed
See
Data Logging
- page102.
cannot view specific data
log records at 5200A
Data log record has older
date and time-stamp than
the 25 most current data log
records available at 5200A
sensor log.
See
Data Logging
- page102.
Troubleshooting
Firmware Updater Troubleshooting
Issue
possible cause
possible solution
Device “timed out” error
incorrect configuration in
5200A communication menu
See Device Timed Out er-
ror - page 112.
incorrect wiring of RS232
connection
See Device Timed Out er-
ror - page 112.
5200A does not go
into downloader mode
when “program device”
is selected at Firmware
Updater
Manually put 5200A in
downloader and retry firm-
ware updater upload.
Menus → System → Down-
loader
See
Force Downloader
-
page
113.
unit does not upload new
code at downloader
unit is not in downloader
mode
See
Downloader
- page 109..
RS232 connection is incor-
rect
See
Wire Communications -
RS232
- page 58.
Feed Timer Troubleshooting
Issue
Understanding how the 5200A calculates sensor reduction
time
Sensor reduction % is
not accurate based on
displayed sensor value(s).
The 5200A displays temperature sensor values to the 10ths
and DO values to 100ths. In reality the 5200A temperature
and DO sensor system values are stored at a much greater
resolution (out to 64 bits). It is the stored values, not the dis-
played values, that are used to calculate the feed timer sensor
reduction.
AquaManager Troubleshooting
Refer to AquaManager online Help for all AquaManager operation and troubleshooting
information.