10
3.2.2
Full Scale Concentration
The voltage output of the analog interface is scaled relative to the calibration concentration. The user can
configure the analog output for each channel to represent 1, 2, 3, or 4 times the calibration concentration.
These outputs are not configured on the PCB. This configuration option is provided via software in the
setup menu. See Section 5.2.6 ANALOG/CONTROL to set this parameter.
3.2.3
Discrete Output Polarity
Three of the discrete control outputs can be configured to obtain either normal or inverted output. These
three outputs are the Black channel pump control, the White channel pump control and the filtrate pump
control output. Switch SW1 on the 2730 PCB contains four dip switches. Switches 1 through 3 control the
outputs. Switch 4 is not used. Figure 7 depicts the relationship between the switch placement and the
polarity of each output.
71#
,QWHUIDFLQJ
714#
6DPSOH
#
,QWHUIDFH
Several methods exist for obtaining the sample for analysis on the YSI 2700. For discrete sampling the
fixed stations (#2 and #3) may be adequate. In many process monitoring and/or control applications the
sample must be transported (pumped) to the analyzer. When configured appropriately the 2700 SELECT
automatically samples a bioreactor, process stream, or other suitable sample source. Since color, turbidity,
optical density and many other physical factors do not affect the YSI enzyme biosensor, filtration and/or
dilution may not be necessary and the 2700/2730 may draw the sample directly. If cell loss is a concern, or
if high cell density is expected, a filtration device (e.g., tangential flow filter) which separates broth and
cells may be installed between the sample source and the 2730 Monitor and Control accessory.
4.1.1
Filtrate Pump
The filtrate pump control output from the YSI 2730 is provided to control a user-supplied external pump
for pumping the sample from its point of origin to the 2700. The timing of this signal is the same as the
purge pump of the YSI 2730 accessory and is setup inside the "Monitor" part of the RunMode menu. This
option is normally not used since the 2730 monitor pump can transport most samples to the external
chamber. If an external filtrate pump is used, the flow rate must be maintained at or above 570
µ
L/minute
in order to guarantee that the sipper does not aspirate air during the sampling process.
SW1 Settings for Output Polarity
Output
Switch
Closed
Open
Black
1
Inverted
Normal
Filtrate
2
Inverted
Normal
White
3
Inverted
Normal
Not Used
4
N/A
N/A
Figure 7 Output Polarity