Your control solution test results will not be included in the average reading
calculations, but can still be recalled. The control solution test result will be
marked with the “CS” symbol on the screen.
The suggested temperature range for the control solution test is
6 - 44 °C (43 - 111 °F).
Do not drip the control solution directly from the bottle onto the sample
entry of the test strip. The reagent on the test strip could be sucked into the
bottle of control solution and could cause the solution’s degeneration.
Doing this could potentially contaminate the meter via the test strip port.
Do not touch the nozzle of the control solution bottle. If you have touched
it, please clean it carefully with water.