The “Battery not suitable” alarm appears if you insert a battery whose voltage is too
high. Confirm by tapping . Remove the unsuitable battery from the battery
compartment and insert a new AAA alkaline battery (LR03). Follow the procedure on
page 53 of this Caregiver Guide.
Battery not suitable
The “Charge internal rechargeable battery” alarm appears if the internal rechargeable
battery of the mylife
™
YpsoPump
®
is depleted on account of a high load. All the insulin
deliveries in progress are cancelled: boluses, temporary basal rate and basal rate.
Confirm by tapping . The internal rechargeable battery is now charged by the alkaline
battery which is indicated by arrows rotating round the hourglass. The charging process
can take up to 20 minutes. When you have confirmed the alarm, any cancelled boluses
and the cancelled temporary basal rate function are indicated with appropriate warnings.
Then the status screen is displayed, but not if the “Charge internal rechargeable battery”
alarm has appeared after the threaded rod has returned or during a self-test after start-
ing the insulin pump from storage status. In these cases, the welcome screen appears.
Charge internal rechargeable battery
Read the User Guide