03002V-90(D)

ANEMOMETER FLANGE BEARING REPLACEMENT

If  anemometer  bearings  become  noisy  or  wind  speed  threshold 

increases above an acceptable level, replace the bearings.  Check 

bearing condition by hanging an ordinary paper clip (0.5 gm) on the 

outside edge of one cup while the instrument is held in a horizontal 

position. The cup should rotate downward.  Failure to rotate due to 

the weight of the paper clip indicates anemometer bearings need 

replacement. Repeat this test at different positions to check full 

bearing rotation. Replace bearings as follows:

1.  REMOVE BEARINGS

 

a)  Loosen set screw on side of cup wheel hub.  Remove cup 

wheel.

 

b)  Remove three screws which hold two sections of main 

 

 

housing.

 

c)  Carefully separate upper and lower housing.  Remove coil 

transducer assembly from upper housing.  Do not disconnect 

from circuit board.

 

d)  Loosen screw and remove ring magnet on end of shaft 

inside  upper housing.

 

e)  Slide shaft and skirt assembly out of both upper and lower 

bearings.

 

f)  Using knife blade under bearing flange, carefully remove 

upper bearing.

 

g)  Using a pencil, gently push out lower bearing from above.

2.  INSTALL NEW BEARINGS

 

a)  Insert new upper bearing.  Use care not to apply excessive 

pressure.

 

b)  Slide cup wheel shaft through upper bearing.

 

c)  Slide lower bearing on shaft inside upper housing.

 

d)  Using ring magnet assembly, push lower bearing into 

 

 

its seat in upper housing.

 

e)  Secure ring magnet to shaft using screw removed in step 

1d.  Use a small amount of sealant on screw to prevent it 

from loosening.

 

f)  Join  two  housing  sections.    Secure  using  three  screws 

removed in step 1b.

 

g)  Place cup wheel on shaft. Tighten set screw on side of hub. 

Do not exceed 80 oz-in torque.

VANE FLANGE BEARING REPLACEMENT

If vane bearings become noisy or if wind direction threshold increases 

above an acceptable level, replace the bearings.  Check bearing 

condition by adding two ordinary paper clips (0.5 gm each) to the 

back edge of the vane fin while the instrument and vane are held 

in a horizontal position. Gently release the vane. It should rotate 

downward. Failure to do so indicates the bearings need replacement. 

Repeat this test at various positions to check full bearing rotation.

Since this procedure is similar to anemometer bearing replacement, 

only the major steps are shown here:

1.  REMOVE BEARINGS

 

(Remove coupling disc - same as ring magnet)

2.  INSTALL NEW BEARINGS

3.  ALIGN VANE (See CALIBRATION section)

EMC COMPLIANCE

This  device  complies  with  Part  15  of  the  FCC  Rules.  Operation 

is subject to the following two conditions: (1) this device may not 

cause  harmful  interference,  and  (2)  this  device  must  accept  any 

interference  received,  including  interference  that  may  cause 

undesired operation.

This  equipment  has  been  tested  and  found  to  comply  with  the 

limits for a Class A digital device, pursuant to part 15 of the FCC 

Rules. These limits are designed to provide reasonable protection 

against harmful interference when the equipment is operated in a 

commercial  environment.  This  equipment  generates,  uses,  and 

can radiate radio frequency energy and, if not installed and used 

in  accordance  with  the  instruction  manual,  may  cause  harmful 

interference to radio communications.  Operation of this equipment 

in a residential area is likely to cause harmful interference in which 

case the user will be required to correct the interference at his own 

expense.

This ISM device complies with Canadian ICES-001.

Cet appareil ISM est conforme à Ia norme NMB-001 du Canada.

EN55011/CISPR 11, Group 1, Class B device.

Class B equipment is suitable for use in domestic establishments 

and  in  establishments  directly  connected  to  a  low  voltage  power 

supply  network  which  supplies  buildings  used  for  domestic 

purposes.

Note:

Wind Sentry sensors with a Voltage Interface Circuit Board may be 

disrupted by radiated RF interference at 80-145 MHz.  To meet EMC 

Compliance,  (2)  YOUNG  18500  ferrite  chokes  must  be  installed 

on  the  cable,  one  near  the  sensor,  and  one  near  the  recording 

instrument.    (Each  choke  must  have  the  cable  pass  through  the 

center hole at least 2 times, creating 1 loop around the outside as 

shown below.)

WARRANTY

This  product  is  warranted  to  be  free  of  defects  in  materials  and 

construction for a period of 12 months from date of initial purchase. 

Liability is limited to repair or replacement of defective item. A copy 

of the warranty policy may be obtained from R. M. Young Company.

CE COMPLIANCE

This  product  has  been  tested  and  complies  with  European  

CE requirements for the EMC Directive.  Please note that shielded 

cable must be used.

Отзывы: