Young WIND MONITOR-AQ 05305 Скачать руководство пользователя страница 3

Page 2

05305-90(T)

disturbed to twice the height of the structure upwind, six times the 

height downwind, and up to twice the height of the structure above 

ground. For some applications it may not be practical or necessary 

to meet these requirements.

FAILURE TO PROPERLY GROUND THE WIND MONITOR

MAY RESULT IN ERRONEOUS SIGNALS

OR TRANSDUCER DAMAGE.

Grounding  the  Wind  Monitor  is  vitally  important.  Without  proper 

grounding  static  electrical  charge  can  build  up  during  certain 

atmospheric conditions and discharge through the transducers. This 

discharge  can  potentially  cause  erroneous  signals  or  transducer 

failure.  To  direct  the  discharge  away  from  the  transducers,  the 

mounting post assembly in which the transducers are mounted is 

made with a special antistatic plastic. Therefore it is very important 

that the mounting post be connected to a good earth ground. There 

are two ways this may be accomplished. First, the Wind Monitor may 

be mounted on a metal pipe which is connected to earth ground. 

The mounting pipe should not be painted where the Wind Monitor is 

mounted. Towers or masts set in concrete should be connected to 

one or more grounding rods. If it is difficult to ground the mounting 

post in this manner an alternative method should be used. Inside the 

junction box the terminal labeled EARTH GND is internally connected 

to the antistatic mounting post. This terminal should be connected to 

an earth ground (Refer to wiring diagram).

Initial installation is most easily done with two people; one to adjust 

the instrument position and the other to observe the indicating device. 

After initial installation, the instrument can be removed and returned 

to its mounting without realigning the vane since the orientation ring 

preserves  the  wind  direction  reference.  Install  the  Wind  Monitor 

following these steps:

1.   MOUNT  WIND  MONITOR 

  a)  Place orientation ring on mounting post.

    Do Not tighten band clamp yet.

  b)  Place Wind Monitor on mounting post.  Do Not tighten band 

clamp yet. 

  

2.  CONNECT  SENSOR  CABLE

  a)  Slide junction box cover up.

  b)  Route cable thru strain relief opening at bottom of junction box. 

Secure cable by tightening packing nut. 

  c)  Connect sensor cable to terminals. See wiring diagram.

  d)  Slide junction box cover down.

3. ALIGN  VANE

  a)  Connect instrument to an indicator.

  b)  Choose a known wind direction reference point on the horizon.

  c)  Sighting  down  instrument  centerline,  point  nose  cone  at 

reference point on horizon.

  d)  While holding vane in position, slowly turn base until indicator 

shows proper value.

  e)  Tighten mounting post band clamp.

  f)  Engage orientation ring indexing pin in notch at  instrument 

base.

  g)  Tighten orientation ring band clamp.

CALIBRATION

The  Wind  Monitor  is  fully  calibrated  before  shipment  and  should 

require no adjustments. Recalibration may be necessary after some 

maintenance operations. Periodic calibration checks are desirable 

and may be necessary where the instrument is used in programs 

which require auditing of sensor performance.

Accurate wind direction calibration requires a Model 18112 Vane 

Angle  Bench  Stand.  Begin  by  connecting  the  instrument  to  a 

signal conditioning circuit which has some method of indicating 

azimuth value. This may be a display which shows azimuth values 

in angular degrees or simply a voltmeter monitoring the output. 

Orient the base with the junction box at 180°. Visually align the 

vane with the crossmarkings and observe the indicator output. If 

the vane position and indicator do not agree within 5°, adjust the 

potentiometer coupling inside the main housing.  Details for making 

this  adjustment  appear  in  the  MAINTENANCE,  potentiometer 

replacement outline, step 7. 

It  is  important  to  note  that,  while  the  sensor  mechanically 

rotates through 360°, the full scale wind direction signal from the 

instrument  occurs  at  355°. The  signal  conditioning  electronics 

must be adjusted accordingly. For example, in a circuit where 0 

to 1.000 VDC represents 0° to 360°, the output must be adjusted 

for 0.986 VDC when the instrument is at 355°. (355°/360° X 1.000 

volts = 0.986 volts)

Wind speed calibration is determined by propeller pitch and the 

output  characteristics  of  the  transducer.  Calibration  formulas 

showing wind speed vs. propeller rpm and output frequency are 

included below. Standard accuracy is ± 0.3 m/s (0.6mph). For 

greater accuracy, the device must be individually calibrated in 

comparison with a wind speed standard. Contact the factory or 

your supplier to schedule a NIST (National Institute of Standards 

& Technology) traceable wind tunnel calibration in our facility.

 

To  calibrate  wind  system  electronics  using  a  signal  from  the 

instrument,  temporarily  remove  the  propeller  and  connect  an 

Anemometer  Drive  (18802  or  equivalent)  to  the  propeller  shaft. 

Apply the appropriate calibration formula to the calibrating motor 

rpm and adjust the electronics for the proper value. For example, 

with the propeller shaft turning at 3600 rpm, adjust an indicator to 

display 18.4 meters per second. (3600 rpm X 0.00512 m/s/rpm = 

18.4 m/s).

CALIBRATION FORMULAS

Model 05305 Wind Monitor-AQ w / 08254 Propeller

WIND SPEED  vs  PROPELLER RPM

m/s 

0.00512 x rpm

knots 

0.00995 x rpm

mph 

0.01145 x rpm

km/h 

0.01843 x rpm

WIND SPEED  vs  OUTPUT 

FREQUENCY

m/s 

0.1024  x Hz

knots 

0.1990 x Hz

mph 

0.2290 x Hz

km/h 

0.3686 x Hz

MAINTENANCE

Given proper care, the Wind Monitor should provide years of service. 

The only components likely to need replacement due to normal wear 

are the precision ball bearings and the wind direction potentiometer. 

Only a qualified instrument technician should perform the replace-

ment. If service facilities are not available, return the instrument to the 

company. Refer to the drawings to become familiar with part names 

and locations. The asterisk * which appears in the following outlines 

is a reminder that maximum torque on all set screws is 80 oz-in.

POTENTIOMETER REPLACEMENT

The potentiometer has a life expectancy of fifty million revolutions.  

As it becomes worn, the element may begin to produce noisy signals 

or become nonlinear. When signal noise or non-linearity becomes 

unacceptable,  replace  the  potentiometer.  Refer  to  exploded  view 

drawing and proceed as follows:

Содержание WIND MONITOR-AQ 05305

Страница 1: ...ICAL INSTRUMENTS INSTRUCTIONS R M YOUNG COMPANY 2801 AERO PARK DRIVE TRAVERSE CITY MICHIGAN 49686 USA TEL 231 946 3980 FAX 231 946 4772 WEB www youngusa com P N 05305 90 REV T102811 WIND MONITOR AQ MO...

Страница 2: ...hen the potentiometer wiper is in the 5 deadband region the output signal is floating and may show varying or unpredictable values To prevent false readings signal conditioning electronics should clam...

Страница 3: ...performance Accurate wind direction calibration requires a Model 18112 Vane Angle Bench Stand Begin by connecting the instrument to a signal conditioning circuit which has some method of indicating a...

Страница 4: ...tiometer coupling properly oriented continue pushing main housing onto vertical shaft bearing rotor until main housing latch locks into position with a click 7 ALIGN VANE a Connect excitation voltage...

Страница 5: ...een tested and shown to comply with European CE requirements for the EMC Directive Please note that shielded cable must be used Declaration of Conformity R M Young Company 2801 Aero Park Drive Travers...

Страница 6: ...Page 5 05305 90 T CABLE WIRING DIAGRAM...

Страница 7: ...Page 6 05305 90 T BEARING REPLACEMENT POTENTIOMETER ADJUSTMENT...

Страница 8: ...Page 7 05305 90 T GENERAL ASSEMBLY REPLACEMENT PARTS...

Страница 9: ......

Страница 10: ...Accessories Model 18802 Anemometer Drive Model18112 Vane Angle BenchStand Model 18310 Propeller Torque Disc Model18331Vane Torque Gauge Model 18301 Vane Alignment Rod Model 18212 Vane Angle Fixture T...

Страница 11: ...Vane Torque Gauge checks vane bearing torque of the Wind Monitor family sensors Slip the fixture over the main housing and make simple yet accurate vane torque measurements Charts relating vane torqu...

Отзывы: