Danger zones M and G
The medical device is not suitable for use in potentially explosive atmospheres or with potentially explosive mixtures
of anaesthetic substances containing oxygen or nitrous oxide according to IEC 60601-1/ANSI/AAMI ES 60601-1.
The medical device is not suitable for use in oxygen-enriched atmospheres.
Zone M
is defined as a »medical environment« and constitutes the part of a room in which potentially explosive
atmospheres may form due to the use of anaesthetics or medical antiseptics and antibacterial soaps;
such atmospheres are typically localized and temporary.
Zone M comprises a truncated pyramid below the operating table which is tilted outwards at a 30° angle.
Zone G
, also known as an »enclosed medical gas system«, does not necessarily include areas enclosed around
all sides, in which explosive mixtures are continuously or temporarily generated, directed or used in small quantities.