13
Ambient
temperature
0
℃
~
37
℃
-20
℃
~
+55
℃
Relative humidity
30%
~
85% (No
condensation)
30%
~
93% (No condensation)
Atmospheric
pressure
860hPa
~
1060hPa
700hPa
~
1060hPa
2.3.6
Probe outline dimensions and weight
Net weight: 50g (R50) (C5-2Fs)
Convex array size: 53mm (L) * 28mm (W) * 38mm (thickness)
Net weight: 70g (R60) (C5-2Ks)
Convex array size: 73mm (L) * 28mm (W) * 38mm (thickness)
Net weight: 30g (L40) (L11-4Ks)
Linear array size: 55mm (L) * 28mm (W) * 30mm (thickness)
Net weight: 20g (L25) (L11-4Gs)
Linear array size: 41mm (L) * 28mm (W) * 30mm (thickness)
WARNING
:
Transportation:
Do not use or store the system outside the specified
environmental conditions.
Working:
1.Make sure the scanner is been held firmly, or it will drop and hurt
the patient.
2.Use the scanner in a dry environment, the operation of
environmental temperature and humidity changes, may lead to
liquid condensation in the circuit board, there is the risk of short
circuit.
3.Do not operate the unit in an environment with flammable or
explosive liquids, vapors or gases such as oxygen or hydrogen.
Do not use the scanner in an environment containing combustible
gas, such as an anesthetic gas, hydrogen, or flammable liquid,
such as ethanol. It may cause an explosion..
A.
If there are flammable substances in the environment, do not
use the scanner.
B. Use the real-time detection environment to detect flammable
substances after the system is turned on. Do not attempt to turn
off the device or unplug the power supply. First empty the air in
the area and ensure a smooth ventilation and then turn off the
power.
4. If the system fails, please do not disassemble the scanner,
please contact the service center or your sales representative.