Yosemite Y505-A Скачать руководство пользователя страница 6

Y505-A Maintenance

1.Maintenance Schedule

Unlike traditional electrical chemical sensors, optical DO sensors require low maintenance.  
There are no need for frequent solution filling and calibrations.

www.yosemitech.com

Maintenance task

Frequency

Sensor cleaning

Every 30 days

Sensor and sensor cap inspection

Every 30 days

Sensor cap replacement

Every one year

Calibration

if required of agency

Calibration based on required schedule

Note

above schedule is only for reference, customers should make on their own judgment  

based on their applications. Yosemite Technologies’ sensor cap may be used longer than  
one year, we recommend to replace the sensor cap per year to ensure performance.

2.Routine Maintenance

DO Probe:

Inspect all sensor connectors for corrosion and replace sensors if necessary.

Inspect sensor o-rings and replace if necessary

After prolonged storage, calibration of the sensor is required.

After use, rinse the sensor body with tap water, dry it, and put on storage cap. Keep DO  

sensor storage cap wet to moisturize DO membrane surface.

DO NOT DRYOUT DO SENSOR CAP!

Inspect the sensor for scratches or cracks, if any are present , replacement of sensor cap

may be needed.

Cable:

Cable’s minimum bending radius is 15mm. The wires inside the cable may break if it is under  
long term stressed condition.

3.Sensor Cap care and Storage

Sensor cap storage

if the senor do not in use, store the sensor in its storage cup. Store the  

sensor cap in water saturated air. If sensor cap has been dried out, please submerge the  
sensor cap in warm (about 40 -50

C) water for 48 hours

Do not expose sensor cap under direct sunlight, keep surface clean, dirty or scratch will  
cause inaccurate reading.

If DO membrane surface is dirty, try to rinse the surface with tap water first. If necessary,  
please gently wipe the surface with lens cleaning tissues that has been moistened with clean  
water. Do not use organic solvents to clean the sensor cap.

Содержание Y505-A

Страница 1: ...Aquaculture Optical Dissolved Oxygen Sensor User Manual Version V1 2 www yosemitech com Yosemite Technologies Co Ltd 2015 All right reserved...

Страница 2: ...used by improper use of this product will not be covered by factory warranty Please keep all documents and if you have any questions please do not hesitate to contact Yosemite Technologies customer se...

Страница 3: ...del Y505 A replacement Cap Y9003 DO monitoring is the most critical water quality parameter in aquaculture Yosemitech ODO series aquaculture optical dissolved oxygen sensor is specially designed for a...

Страница 4: ...Y505 A Y505 A www yosemitech com...

Страница 5: ...protocols Installation thread M22 1 5 mm Power Requirements DC 6 12V 5 current 50mA Temperature Sensor Type NTC Sensor OD 23mm Sensor Length 188 3 mm Cable Length 10m standard 5m 15m and 30m optional...

Страница 6: ...rage calibration of the sensor isrequired After use rinse the sensor body with tap water dry it and put on storage cap KeepDO sensor storage cap wet to moisturize DO membranesurface DO NOT DRYOUT DO S...

Страница 7: ...2 Power Supply Requirements Power Supply DC 6 12V 5 Current 50mA 3 Sensor Cable 4 wire AWG 24 or AWG 26 shielding wire OD 5mm 1 Red Power VCC 2 White 485 Date_B 485_B 3 Green 485 Date_A 485_A 4 Black...

Страница 8: ...ntroller Cable failure Contact customerservice Bad sensor cap or loose sensor cap Reinstall sensor cap orreplace Unstable DO reading Dirty sensor cap Clean sensor cap Damaged sensor cap Replace sensor...

Страница 9: ...on improper or unreasonable use or maintenance misuse abuse accident neglect exposure to fire improper packing and shipping lightning power surges or other acts of nature This limited warranty does no...

Страница 10: ...CO LTD Mailing address Bldg NW 06 Ste 502 99 JinJi Lake Ave Suzhou Industrial Park China Zip code 215123 Tel 86 512 87663987 86 512 87663997 86 512 65970621 Fax 86 512 87663911 E mail sales yosemitech...

Отзывы: