Installation
Part List
Item
Number
Note
Y504-A Sensor
1
Cable
1
10m
protective cover
1
Rubber protective Cap
1
Before use
1) Take off the protect cap
: Please take off the protect cap of Optical dissolved
oxygen sensor , before installation and keep them properly for future use.Meanwhile,
tighten the protective cover.
Sensor Installation
1) Wiring and power supply:
•
The female and male connector of sensor cable should be screwed tightly to
avoid moisture incursion
•
Do not use the sensor cable to pull the sensor! It is required to install sensor in a
secure and stable mounting bracket.
•
Make sure power supply voltage is correct before power on.
2) Sensor installation:
•
It is recommended to install the sensor vertically with electrodes facing down.
•
Considering water level change, the sensor should be installed 30cm below
water level. The sensor should not be installed no more than 2m below water
surface for maintenance purpose.
•
The sensor must be securely installed to avoid damage caused by water flow and
other things.