York Survey Supply 354920 Скачать руководство пользователя страница 2

©York Survey Supply Centre 2018

Ref:.. \operat98\instructions 18\354920.qxp 05-01-18

Protimeter Hygromaster

with Humidity Box or Humidity Sleeves

Moisture Related Damage in Floors

Excessive moisture in concrete floor slabs and screeds can lead to the following types of failure or

damage:

•  Adhesion failure between floor covering and concrete slab. When the critical moisture level for 

a given covering/adhesive/concrete is passed, bond strength drops sharply.

•  Vinyl floor covering movement. Dampness in concrete may aggravate alkali related 

dimensional changes in sheet vinyl.

•  Emissions of harmful substances from floor covering materials. Examples include the emission 

of ammonia from some screeds, the emission of formaldehyde from chipboard and the 

degradation of certain plasticisers in sheet vinyl.

•  Excessive expansion of solid wood floor coverings and the decay and degradation of wood 

based floorboards.

These problems tend to occur when the moisture content of the concrete or screed exceeds the

critical moisture content of the materials that are in contact with it. A material’s critical moisture

content can be expressed as an equilibrium relative humidity (ERH) value or a moisture content

(%mc) value. ERH values are particularly suitable for this purpose because they convey the 

moisture condition of the material in question - its degree of wetness. Whilst the critical moisture

content values vary for different materials, the problems outlined above are avoided when 

substrates have an ERH below 75%.

Humidity and Relative Humidity (%rh)

Humidity denotes the presence of water vapour - the gaseous form of water - in air and other

gases. The expression relative humidity (rh%) of air expresses the degree of saturation with water

vapour at a given temperature. Saturated air - having 100%rh - contains its full capacity of water

vapour. Relative humidity measurements are particularly appropriate for assessing the moisture

condition of environments. Similarly, equilibrium relative humidity (ERH) measurements are very

useful for assessing if materials are in a wet or dry condition. ERH values of materials are 

obtained by measuring the %rh of a pocket of air deemed to be in moisture equilibrium with the

material under investigation.

Guidance for Measuring Moisture in the ERH of Screeds and Slabs

The GE Protimeter Hygromaster L and GE Protimeter MMS instruments may be used to measure

the ERH of screeds and floor slabs when they are used with a GE Protimeter extension lead and a

GE Protimeter humidity box or GE Protimeter humidity sleeves. Please refer to the GE Protimeter

Hygromaster L and MMS user instructions for specific information on using these instruments.

1. ERH Test Procedure - using GE Protimeter Humidity Sleeves

i.  Drill 16mm (5/8”) diameter clearance holes to a nominal depth of 50mm (2”) at the required 

points of measurement (POMs).

ii.  Position a plastic membrane over the hole and push a GE Protimeter humidity sleeve through 

the membrane and into the hole. Ensure that the flange of the humidity sleeve is flush with the 

floor surface and that the humidity sleeve cap is firmly in position, to make an airtight seal.

iii.  Leave for at least 24 hours to ensure moisture equilibrium conditions are reached within the 

humidity sleeve. If in use, switch off any accelerated drying apparatus at least 4 days before 

taking final readings.

iv.  The ERH reading is made with the Hygrostick sensor supplied with the GE Protimeter 

Hygromaster L instrument. Hygrosticks are fitted with a soft conical sealing sleeve. Ensure that 

the tip of the Hygrostick protrudes fully from the narrow end of the sealing sleeve.

v.  Remove the humidity sleeve cap and push the Hygrostick assembly into the humidity sleeve 

and twist firmly to seal. Ideally, leave the Hygrostick in the hole for at least 30 minutes before 

taking the first reading.

vi.  Connect the Hygrostick to the Hygrometer L instrument with the GE Protimeter extension lead.

vii. Measure the %rh in the hole; this is the ERH value of the floor slab.

2. ERH Test Procedure - using GE Protimeter Humidity Box

This test procedure follows the guidelines given in the British Standards BS 5325 and BS 8203. 

If full BS procedure compliance is required, please refer to these standards for comprehensive 

details.

i.  Place the humidity box on the surface at the required point of measurement (POM) and put a 

brick or similar object on top to ensure it is not inadvertently moved. Ensure the plug is in the 

hole in the side of the box. If the surface is smooth, no additional sealing is required around the 

rim of the box. If the surface has a rough texture, seal the box-surface interface with an inert 

sealant such as plasticine.

ii.  If in use, switch off any accelerated drying apparatus at least 4 days before taking readings. 

Thereafter, for un-bonded screeds of thickness 50 - 75mm (2 - 3”) where a damp-proof 

membrane has been placed between the base and screed, leave the humidity box in position 

for at least 24 hours to ensure moisture equilibrium is reached within the box before taking 

readings. For bonded screeds and direct finish base slabs allow a minimum of 72 hours to 

pass before taking readings. Note that it may take many weeks for equilibrium conditions to be 

reached within a humidity box on the surface when the slab thickness exceeds 200mm (8”) 

and/or it has a power floated finish.

iii.  The ERH reading is made with the Hygrostick sensor supplied with the GE Protimeter 

Hygromaster L instrument. Hygrosticks are fitted with a soft conical sealing sleeve. Ensure that 

the tip of the Hygrostick protrudes fully from the narrow end of the sealing sleeve.

iv.  Remove the humidity box plug and push the Hygrostick assembly into the hole and twist firmly 

to seal. Ideally, leave the Hygrostick in the humidity box for at least 30 minutes before taking 

the first reading.

v.  Connect the Hygrostick to the Hygrometer L instrument with the GE Protimeter extension lead.

vi.  Measure the %rh in the humidity box; assuming equilibrium conditions have been reached, this 

is the ERH value of the screed or slab.

Отзывы: