![YOKOGAWA XS550 Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/yokogawa/xs550/xs550_user-manual_3518743034.webp)
8-3
<General Specifications>
IM 01W06F02-01EN
Table 8.2 Effects of Ambient Temperature
Sensor Type
Temperature Effects per 1.0°C
Change in Ambient
Temperature
Measurement Range
T/C
B
0.092°C – (0.032% of (t –100))
100°C ≤ t < 300°C
0.025°C – (0.0010% of (t –300))
300°C ≤ t < 1000°C
0.029°C
1000°C ≤ t < 1820°C
E
0.00099°C – (0.0040% of t)
–200°C ≤ t < 0°C
0.00099°C + (0.0019% of t)
0°C ≤ t < 1000°C
J
0.0012°C – (0.0041% of t)
–200°C ≤ t < 0°C
0.0012°C + (0.0019% of t)
0°C ≤ t < 1200°C
K
0.0015°C – (0.0048% of t)
–200°C ≤ t < 0°C
0.0015°C + (0.0023% of t)
0°C ≤ t < 1372°C
N
0.0023°C – (0.0056% of t)
–200°C ≤ t < 0°C
0.0023°C + (0.0019% of t)
0°C ≤ t < 1300°C
R, S
0.013°C – (0.020% of t)
–50°C ≤ t < 0°C
0.013°C + (0.0021% of t)
0°C ≤ t < 100°C
0.0093°C + (0.0012% of t)
100°C ≤ t < 600°C
0.0030°C + (0.0021% of t)
600°C ≤ t < 1768°C
T
0.0016°C – (0.0044% of t)
–200°C ≤ t < 0°C
0.0016°C + (0.0015% of t)
0°C ≤ t < 400°C
C
0.0046°C + (0.0014% of t)
0°C ≤ t < 400°C
0.00013°C + (0.0024% of t)
400°C ≤ t < 1400°C
–0.023°C + (0.0041% of t)
1400°C ≤ t < 2000°C
–0.11°C + (0.0085% of t)
2000°C ≤ t < 2315°C
Note1: The “t” on Table 2 means the value of the reading in°C.