![YOKOGAWA WT300 Скачать руководство пользователя страница 65](http://html.mh-extra.com/html/yokogawa/wt300/wt300_user-manual_911408065.webp)
2-35
IM WT310-02EN
Making Preparations for Measurements
1
2
3
4
5
6
7
App
Index
Connecting to the WT332/WT333
In the wiring examples that follow, the WT332/WT333 input elements, voltage input terminals, and
current input terminals are simplified as shown in the following figure.
VOLTAGE
±
±
CURRENT
EXT
EXT
C
±
± V
Input element
Voltage input
terminals
Current input
terminals
WT332/WT333
Also, the wiring example is for when a shunt-type current sensor is connected. When connecting
a clamp-type current sensor that outputs voltage, substitute the shunt-type current sensor with the
clamp-type current sensor.
Shunt-type current sensor
±
I
OUT L
OUT H
Input element
Input element
Clamp-type current sensor
that outputs voltage
EXT
C
±
± V
EXT
C
±
± V
Wiring Example of a Single-Phase, Two-Wire System (1P2W) with
a Shunt-Type Current Sensor
The following wiring example shows how to configure the wiring to connect to input element 1. To
configure the wiring for other input elements, substitute the numbers in the figures with the appropriate
element numbers.
SOURCE
LOAD
Earth side
OUT L
OUT H
Element 1
EXT
C
±
± V
Shunt-type current sensor
2.10 Wiring the Circuit under Measurement When Using Current Sensors