17-9
IM WT1801-01EN
Single Shot (Single-shot storage mode)
Time
STORE START
key
State
Start
STORE START
on
Cmpl
Storage count
STORE STOP
on
STORE RESET
key
Storage Operations Performed When You Press STORE STOP After Storage Has Been
Started
Restarting Storage
When you press STORE START, STORE STOP, and then STORE START
Time
State
Start
STORE START
on
Stop
STORE START
key
STORE STOP
blinking
STORE START
key
STORE STOP
key
Start
STORE START
on
Resetting Storage
When you press STORE START, STORE STOP, and then STORE RESET
Time
State
Start
STORE START
on
Stop
STORE RESET
key
STORE STOP
blinking
STORE START
key
STORE STOP
key
Storage Operation When the Storage Interval Is Not an Integer Multiple of the Data
Update Interval
Here we will use the case when the data update interval is 5 s and the storage interval is 7 s as an example.
How storage is performed varies as indicated below depending on the storage mode.
When the Storage Mode Is Not Set to Integ Sync
Storage is performed according to the storage interval.
Time (s)
Data
update
Storage
Time (s) 0 5 10 15 20 25 30 35
0 7 14 21 28 35
When the Storage Mode Is Integ Sync and the Integration Mode Is Not Continuous or R-Continuous
Storage is performed at the first data update after the storage interval passes.
Time (s)
Data
update
Storage
Time (s) 0 5 10 15 20 25 30 35
0 7 14 21 28 35
17 Data Storage