7-7
IM 253401-01E
7
Integration
7.4
Precautions Regarding Use of Integrator
Function
Relation between Integration Hold and the START/STOP key
When the HOLD key is pressed, the display and communication output of the integrated results
is being held while integration continues. The relation between this hold function and the
START/STOP key is as follows.
• Even when starting integration while the hold function is on, the display and communication
output will remain unchanged. Only canceling the hold function or activating a trigger
(pressing the SHIFT key followed by the HOLD (TRIG) key) will result in displaying or
outputting the integrated results of the time of cancellation.
ON
HOLD
OFF
Displayed
value
Integration
preset time
(Dotted line shows integrated value)
START
STOP RESET
• Even when stopping integration while the hold function is on, the displayed integrated value
will remain unchanged. However, as soon the hold function is turned off or a trigger is
activated, the integrated results of the time when integration was stopped will be displayed or
output.
Integration
preset time
(Dotted line shows integrated value)
TRIG
OFF
ON
HOLD
ON
ON
ON
Displayed
value
START
STOP
RESET
Relation between Integration Reset and the START/STOP key
The relation between integration reset and the start/stop key is as follows.
Integrated
value
Integration
preset time
Interrupt
Start
Restart
Interrupt
Reset
Restart
Auto stop
Reset
Preset time for integration
START
STOP
START
STOP
RESET
START
RESET