11.LIST OF PARAMETERS
No.
Item
Display
Remark
No.
Item
Display
Remark
01
Model
MODEL
02
Tag no.
TAG NO
03
Self-check result
SELF CHK
Display items
A
Display1
DISPLAY1
B
Display2
DISPLAY2
A01
Input value
INPUT1
*1
B01
Input value
INPUT1
A05 Output value 1
OUTPUT1
B05
Output value 1
OUTPUT1
A06 Output value 2
OUTPUT2
B06
Output value 2
OUTPUT2
A07 Alarm-1 status
ALM1 STATUS
B07
Alarm-1 status
ALM1 STATUS
A08 Alarm-2 status
ALM2 STATUS
B08
Alarm-2 status
ALM2 STATUS
A31 Integrating counter 1
COUNTER1
B31
Integrating counter 1
COUNTER1
A32 Integrating counter 2
COUNTER2
B32
Integrating counter 2
COUNTER2
A33 Integrating counter 3
COUNTER3
B33
Integrating counter 3
COUNTER3
A34 Integrating counter 4
COUNTER4
B34
Integrating counter 4
COUNTER4
A54 Status
STATUS
*2
B60
Self-check result
SELF CHK
A56 Rev. no REV NO *1: Instantaneous frequency is displayed evenif conversion mode is pulse integration.
A58 MENU REV
MENU REV
*2: The Status is displayed for service personnel to see history records.
A60 Self-check result
SELF CHK
Setting items
D Setting (I/O)
SET(I/O)
E
Setting (alarm output)
SET(ALM)
D01 Tag no. 1
TAG NO.1
E03
Alarm-1 setting
SET POINT1
D02 Tag no. 2
TAG NO.2
E04
Alarm-2 setting
SET POINT2
D03 Comment 1
COMMENT1
E05
Direction of alarm-1 action ALM1 ACTION
D04 Comment 2
COMMENT2
E06
Direction of alarm-2 action ALM2 ACTION
D10 Range unit
UNIT
E09
Alarm-1 hysteresis
HYSTERESIS1
D13 Low input cut point LOW CUT
E10
Alarm-2 hysteresis
HYSTERESIS2
D19 Conversion mode
SELECT MODE
E11
Alarm-1 ON delay setting ON DELAY1
D20 Sample mode
SAMPLE MODE
E12
Alarm-2 ON delay setting ON DELAY2
D21 Sample time
SAMPLE TIME
E13
Alarm-1 OFF delay setting OFF DELAY1
D22 Input low range
INPUT1 L_RNG
E14
Alarm-2 OFF delay setting OFF DELAY2
D23 Input high range
INPUT1 H_RNG
E15
Direction of alarm-1 relay action RL1 ACTION
D38 Direction of output-1 action
OUT1 DR
E16
Direction of alarm-2 relay action RL2 ACTION
D39 Direction of output-2 action
OUT2 DR
E60
Self-check result
SELF CHK
D50 Input filter
INPUT FILTER
F
Setting (communication)
SET(COM)
D60 Self-check result
SELF CHK
F01
Communication protocol
PROTOCOL
F02
Address
ADDRESS
F03
Baud rate
BAUD RATE
F04
Parity
PARITY
There are items not displayed depending on what output-2 is specified.
F05
Data length DATA LEN
F06
Stop bit
STOP BIT
F07
Decimal point position of input INPUT DEC PT
F60
Self-check result
SELF CHK
Adjusting items
Test items
P Adjustment
ADJUST1
Q
Test
TEST
P12 0% adjustment of output-1
OUT1 0%
Q02
Forced output-1
OUT1 TEST
P13 100% adjustment of output-1
OUT1 100%
Q03
Forced output-2
OUT2 TEST
P14 0% adjustment of output-2
OUT2 0%
Q04
Forced output (alarm)
ALM1 TEST
P15 100% adjustment of output-2
OUT2 100%
Q05
Forced output (alarm)
ALM2 TEST
P60 Self-check result
SELF CHK
Q60
Self-check result
SELF CHK
12.MAINTENANCE
The product starts running immediately when the power is turned on;
however, it needs 10 to 15 minutes of warm-up before it meets the
specified performance.
For cleaning the instrument, use a soft and dry cloth.
12.1 Calibration Apparatus
●
A pulse generator (Yokogawa FG300 or the equivalent): 1
●
A volt meter (Yokogawa 7562 or the equivalent) : 1
●
A counter (Hewllet-Packard Japan 5334B or the equivalent): 1
●
A precision resistor (250
⍀
±0.01% 1W): 1
12.2 Calibration Procedure
Connect the instruments as shown in Fig.12.1.
Produce input signals equivalent to 0, 25, 50, 75 and 100% of the in-
put span from the pulse generator to the converter. Then, check that
the converter's output signal shows voltages corresponding to 0, 25,
50, 75, and 100% of the input span within the rated accuracy range.
If the output signal is out of the rated accuracy range, adjust the out-
put signal level using the Handy Terminal (JHT200). For adjust-
ment, refer to Instruction Manual for VJ77 Parameters Setting Tools
(IM 77J01J77-01) or for Handy Terminal (IM JF82-01).
10
11
3
2
1
4
5
6
7
8
9
Input
Output-2
Output-1
Power supply
9
7
4
3
+
-
Pulse
Generator
R
+
-
5
2
Voltmeter
Voltmeter
R
R:For current output using 250
Ω
precision resistor
R:For current output using 250
Ω
precision resistor
+
-
11
10
8
L+
N-
GND
Fig. 12.1