<3. INSTALLATION, PIPING AND WIRING>
3-13
IM 12T03A01-02E
3.4.1
Wiring between Detector and Converter
This wiring is to be carried out with special cable of specified length. The cable is terminated as
illustrated in Figure 3.22. Pay attention not to stain or wet terminated.
GRAY
YELLOW
BROWN
RED
RED
WHITE
WHITE
BLUE
BLUE
GREEN
GREEN
BROWN
YELLOW
GRAY
Density detector
Density converter
B4
B5
A4
A2
A3
A5
A6
B4
B5
A2
A3
A4
A5
A6
Figure 3.22
DM8W Cable
< Connection to Density Detector >
Remove the cover of the terminal box, using the spanner in accessories, and insert the special
cable into the cable inlet. Connect each conductor to the respective terminals.
A flexible fitting is used at the inlet of the density detector *1 and the cable from the detector runs
though metal conduit to the duct. For Model VD6DF explosion-proof detector, wiring should be
made complying with the Recommended Practice.
The explosionproof flexible fitting for wiring should be used at the cable inlet (G3/4 female) of the detector.
*1:
The flexible fitting is not inevitable. The Detector not moved for usual check or maintenance does not require the flexible tube.
Figure 3.23 Wiring to Density Detector Terminals
Figure 3.24
Spanner for Open/closing
Terminal
Box Cover
Cable
Terminal box case
Lock nut
G3/4 screw (TIIS explosionproof)
3/4NPSM (FM explosionproof)
Explosionproof flaxible fitting
Be sure to screw in
at least 12 mm (5 threads)
Figure 3.25
Explosion-proof Flexible Fitting
5th Edition : Dec. 11, 2015-00
Содержание vigilantplant DM8
Страница 1: ...User s Manual R Model DM8 Vibration Type Density Meter IM 12T03A01 02E IM 12T03A01 02E 5th Edition ...
Страница 5: ...Blank Page ...
Страница 9: ...Blank Page ...
Страница 25: ...Blank Page ...
Страница 41: ...Blank Page ...
Страница 57: ...Blank Page ...
Страница 63: ...Blank Page ...