background image

Instruction
Manual

IM TX10-02E

August 2001, 2nd Edition

Thank you for purchasing this instrument.

For proper use, please read through this instruction manual prior to operation.

TX10-02, -03
Digital Thermometers

This instruction manual describes the Model TX10-03, which has more functions than the Model TX10-02. 
However, with the exception of the            key, operation of the Model TX10-02 is performed in the same 

;

;

;

;

1. Cautionary Notes for Safe Use of the Product

The following safety symbols are used on this instrument and instruction manual to ensure
safe use.

WARNING

This symbol found on the back of the instrument, calls attention to a potential
danger or hazard capable of resulting in a serious injury or loss of life, and the
operator must refer to this instruction manual to avoid the danger or hazard.

 

WARNING

:

This symbol in this manual indicates a potential danger or hazard capable of resulting in
a serious injury or loss of life.

 

CAUTION

:

This symbol in this manual indicates a potential danger or hazard capable of resulting in
severe but not irreversible injury or damage.  In some instances, the hazards may be
those associated with WARNING symbols but are of significantly less magnitude.

 

NOTE

:

This symbol draws attention to information that is essential for understanding how to
operate the instrument and/or instrument features.

TIP

:

Gives information that complements the present topic.

To avoid accidents, such as an electrical shock, which may lead to injury or loss of life to the
operator, or damage to the equipment, be sure to observe the following precautions.

 WARNING

Do not use the instrument to measure where there is the risk of electrical shock. Also, when using the 2-
channel model, always keep the probe-to-probe potential below 1 V.

When handling a needle-type probe, be careful not to point the tip in the direction of any person as doing so
may result in an injury.

After measuring anything with a high temperature, do not touch the metal part of the measuring probe as this
may result in a burn.

 CAUTION

When the instrument is stored for a long period of time, be sure to remove the batteries. Not doing so may
result in a failure or malfunctioning of the instrument due to a leakage in battery liquid.

Do not use deteriorated or damaged probes, as doing so will effect the precision of measurements.

Before you detach the probe from the instrument, always remove the probe from the object being measured.

When using a needle-type probe, do not insert the metal part of the measuring probe more than half its the
length into the object to be measured, as doing so may result in a burn or damage to the probe due to the
handle of the probe heating up.

Keep the handle of the probe and the cable that connects the probe to the instrument within a temperature
range of -20°C to 50°C, as they are less resistant to heat than the metal part of the measuring probe.

Regarding Disassembling and Modification

Do not disassemble or modify the instrument in any way.  The instrument should only be disassembled by a
service engineer of Yokogawa.

2. Functional Description

Thermocouple type select key (Type K/J/E/T)

Maximum and minimum
value record key
Data record key

Relative display select key
Simplified correction mode key

Data call-up key

Power ON/OFF key

Data hold key

Resolution select key

Channel A

Channel B

Display

Maximum, minimum,
and stored data read key

Input channel select key

AUTO OFF

MAX

HOLD RCD REL

MEM
ADJ

A

MIN

B

ETKJ

Battery alarm

Relative display mode

Selected thermocouple type

Auto power-off status

Unit

Displayed in simplified
correction mode.

Displayed when stored
data is shown.

Displayed when max. or 
min. value is shown.

Data hold status

Maximum-minimum
record mode

Selected channel

Select

channel

;

;

;;

;;

(TX10-03

only)

To select the input channel from channel A, B, or A-B:

Press this key.
With each press, the display changes between A, B, and A-B, allowing
you to select the input channel for measurement.

 NOTE

:

• Channel switching is disabled during the RCD mode.
• By switching the channels when data is being read, you can

display the data of each channel: maximum and minimum values
and measured data saved using the Memory-in function.

A

B

A

B

Hold

Release hold

Use this key to hold the measured value on LCD.

To hold, press this key.

To cancel holding, press this key again.

HOLD

Select

resolution

To select a display resolution:

Press this key.
Every time this key is pressed, the display resolution alternates
between 0.1°C and 1°C (within the range of -200.0°C to 199.9°C).
If 0.1°C is selected, a resolution of 0.1°C is achieved when the display
value is 199.9°C or below; when it goes over 200°C, the resolution
changes to 1°C, and the number of display digits is switched automati-
cally. When “chA-B” is selected, the display is switched to the input
value of channel A or B, whichever is greater.

TIP

:

• The resolution setting is maintained even after the power is turned off.

The next time the power is turned on, the instrument starts up with the
setting that was set before the power was turned off.

• Factory-set resolution is 0.1°C.

Relative

display

Cancel

relative
display

To select relative display:

Press this key to select a relative display.
The display shows the difference (relative value) of a subsequent
measured value (Dx) from the reading given as a reference value (D1)
when this key was pressed.
Relative display value = Dx–D1
The reference value affects measurements on all channels (channel A,
channel B and channel A-B).  Any value among MAX, MIN and MEM
can also be used as the reference value.

To cancel relative display, press this key again.

REL

Set

correction

value

To set the correction value for simplified correction:

The current temperature reading can be corrected to a value based on the
reference thermometer.  By thus eliminating probe-specific errors, you can
take precision measurements over a frequently used temperature range.

 NOTE

: When you change the measurement probe, the correction

value must be set again regardless of the thermocouple type.

Press and hold this key for 3 seconds.
The ADJ on the display flashes, indicating that the instrument is ready
for setting.

While measuring the reference temperature, correct the display value to
an optional value using the 

 and 

 keys. (The correction range is

±20°C of the measured value. Resolution of the correction value is
determined according to the resolution setting.)

TIP

: Holding down the 

 or 

 key increases the speed with which

the display value changes.

Press the 

 key to accept the setting.

ADJ flashes.

Key

Operation and description

Display

Power on

Power off

Press this key to turn on the power.

• The auto power-off function turns off the power automatically about 10

minutes after the last key operation of any key other than the 
key.

• To disable the auto power-off function, simultaneously press the

 key and 

 key. With this function disabled, the

instrument will stay powered continuously.

To turn off the power, press this key for 1 second or longer.

After the power
has been turned
on, all the
elements of the
display light up.

AUTO OFF

Select

thermo-

couple type

+

To select a thermocouple type from K, J, E, and T:

If the power is already turned on, turn it off.

Press and hold down the 

 key while pressing the 

 key

to turn on the power. The thermocouple type and unit on the display
flash, indicating that the instrument is ready for changing thermocouple
type.

Every time the 

 key is pressed the thermocouple type and unit

(°C/°F) changes.

Press the 

 key to accept the setting.

TIP

: The factory-set default is type K.

K, J, E, T

Simplified

correction

mode

Cancel

simplified

correction

mode

Use this key to enter and exit the simplified correction mode:

To enter, press this key for 1 to 2 seconds.

To exit, press this key for 1 to 2 seconds again.

ADJ

Record

maximum

and minimum

values

Cancel

recording

Recording of the maximum and minimum values:

Detection and recording of the max./min. values starts when this key is
pressed and continues until this key is pressed again or the power is
turned off.

To start recording, press this key.

 NOTE

: Enable the RCD mode after selecting the channel.

To stop the recording, press this key again.

• Measurement on channel A-B is done as described below.

Press the 

 key to read the minimum and maximum values.  With

the 

;

;

;;

;;

 key, select channel A or B.  The display shows the MIN

and MAX values for each channel.

°

C

t

Channel-A MIN value

Channel-A MAX value

Channel A-B MAX value

Channel A-B MIN value

chA

chB

Channel-B MAX value

Channel-B MIN value

(Start of MIN/MAX recording)

(End of MIN/MAX recording)

RCD

Using the

unit °F

To users who use Fahrenheit as the temperature unit:

To use the unit °F, follow the instructions below.

When turning on the power, keep the 

 key held down.

• Verify that the display shows the °F symbol, and then release the key.
• After completing this step, you can switch between °C and °F

whenever you set the type of thermocouple.

• By following the same procedure as above, you can revert to the

condition in which only the unit °C is available.

°F

   KIM3E-2002.7

  

Printed in Japan

YOKOGAWA M&C CORPORATION

International Sales Dept.
2-9-32 Nakacho, Musashino-shi, Tokyo, 180-8750 Japan
Phone: 81-422-52-5716  Facsimile: 81-422-55-8954

YOKOGAWA CORPORATION OF AMERICA (U.S.A.)
Phone: 1-770-253-7000  Facsimile: 1-770-251-2088

YOKOGAWA EUROPE B. V. (THE NETHERLANDS)
Phone: 31-334-64-1611  Facsimile: 31-334-64-1610

YOKOGAWA AMERICA DO SUL S. A. (BRAZIL)
Phone: 55-11-5681-2400  Facsimile: 55-11-5681-1274

YOKOGAWA ENGINEERING ASIA PTE. LTD. (SINGAPORE)
Phone: 65-6241-9933  Facsimile: 65-6241-2606

YOKOGAWA MEASURING INSTRUMENTS KOREA CORPORATION (KOREA)
Phone: 82-2-551-0660 to -0664  Facsimile: 82-2-551-0665

YOKOGAWA AUSTRALIA PTY. LTD. (AUSTRALIA)
Phone: 61-2-9805-0699  Facsimile: 61-2-9888-1844

YOKOGAWA BLUE STAR LTD. (INDIA)
Phone: 91-80-227-1513  Facsimile: 91-80-227-4270

LTD. YOKOGAWA ELECTRIC (RUSSIAN FEDERATION)
Phone: 7-095-737-7868  Facsimile: 7-095-737-7869

Отзывы: