<6. PROCEDURES FOR CONVERTER OPERATION>
6-21
IM 12E04A02-02E
3rd Edition : May. 31, 2010-00
Table 6.9 Averaging Coef
fi
cient and 90 % Response Time
Averaging coef
fi
cient
Standard for 90 % response time
10
Approx. 8 sec.
20
Approx. 22 sec.
30
Approx. 15 sec.
40
Approx. 30 sec.
50
Approx. 38 sec.
100
Approx. 75 sec.
The following is the procedure for the setting operation.
(Example of display)
1. Select FUNCTION "4" in the <PROGRAM1>
=> 00030
mode.
2. Press the [>] key.
=> 00030
The most signi
fi
cant digit
fl
ashes.
3. Select a digit using the [>] key and set it to the
=> 00040
desired value using the [^] or [v] key.
The
fl
ashing digit moves to the right
and when the least signi
fi
cant digit is
reached, it returns to the most
signi
fi
cant digit.
4. Press the [ENT] key to enter the value.
=> 00040
The entire display
fl
ashes once.
In addition, in step 4, if the set value is beyond the permissible range (1 to 255), "not" is
displayed, Reset the message by pressing the [>] key and repeat operations from step 2.
Note: If calibration is performed using the calibration plate in FUNCTION "4" in the
<MAINT.> mode, the averaging coef
fi
cient automatically changes to 10 regardless of
this setting.
(3) Setting Output Value If an Error Occurs (FUNCTION "5")
If "preset" (_ _3) is selected in the output mode setting (FUNCTION "5" in the <PROGRAM1>
mode) when an error occurs, the analog output when the error occurs is
fi
xed at this setting
value. Display and set the setpoint in the ratio (%) to the analog output span. The following is an
example.
(Example)
For a setpoint at "50 %",
When 4 to 20 mA output:
The output value holds at 12 mA when an error occurs.
When 1 to 5 V output:
The output value holds at 3 V when an error occurs.
The following is the setting key operation.
(Example of display)
1. Select FUNCTION "5" in the <PROGRAM1>
=> 00000 %
mode.
2. Press the [>] key.
=> 00000 %
The most signi
fi
cant digit
fl
ashes.
3. Select a digit with the [>] key and set it to the
=> 00050 %
desired value using the [^] or [v] key.
The
fl
ashing digit moves to the right
and when the least signi
fi
cant digit is
reached, it returns to the most
signi
fi
cant digit.
4. Press the [ENT] key to enter the set value.
=> 00050 %
The entire display
fl
ashes once.