3-38
IM 704310-12E
INPut:BCHannel?
Function
Queries all settings related to channel B.
Syntax
:INPut:BCHannel?
Example
:INPUT:BCHANNEL?
→
:INPUT:BCHANNEL:IMPEDANCE I1M;
COUPLING DC;TRIGGER:MODE MANUAL;
LEVEL 2.500E+00
INPut:BCHannel:COUPling
Function
Sets the coupling of channel B or queries the
current setting.
Syntax
:INPut:BCHannel:COUPling {AC|DC}
:INPut:BCHannel:COUPling?
Example
:INPUT:BCHANNEL:COUPLING DC
:INPUT:BCHANNEL:COUPLING?
→
:INPUT:BCHANNEL:COUPLING DC
INPut:BCHannel:IMPedance
Function
Sets the input impedance of channel B or
queries the current setting.
Syntax
:INPut:BCHannel:IMPedance {I50|I1M}
:INPut:BCHannel:IMPedance?
Example
:INPUT:BCHANNEL:IMPEDANCE I1M
:INPUT:BCHANNEL:IMPEDANCE?
→
:INPUT:BCHANNEL:IMPEDANCE I1M
INPut:BCHannel:TRIGger?
Function
Queries all settings related to the trigger of
channel B.
Syntax
:INPut:BCHannel:TRIGger?
Example
:INPUT:BCHANNEL:TRIGGER?
→
:INPUT:BCHANNEL:TRIGGER:MODE MANUAL;
LEVEL 2.500E+00
INPut:BCHannel:TRIGger:LEVel
Function
Sets the trigger level of channel B or queries
the current setting.
Syntax
:INPut:BCHannel:TRIGger:LEVel <voltage>
:INPut:BCHannel:TRIGger:LEVel?
During trigger manual mode
<level>=-5V to +5V (in 1 mV steps)
During trigger single/repeat auto
<level>=0 to 100% (in 1% steps)
Example
:INPUT:BCHANNEL:TRIGGER:LEVEL 2.5V
:INPUT:BCHANNEL:TRIGGER:LEVEL?
→
:INPUT:BCHANNEL:TRIGGER:
LEVEL 2.500E+00
INPut:BCHannel:TRIGger:MODE
Function
Sets the trigger mode of channel B or queries
the current setting.
Syntax
:INPut:BCHannel:TRIGger:MODE {MANual|
SINGleauto|REPeatauto}
:INPut:BCHannel:TRIGger:MODE?
Example
:INPUT:BCHANNEL:TRIGGER:MODE MANUAL
:INPUT:BCHANNEL:TRIGGER:MODE?
→
:INPUT:BCHANNEL:TRIGGER:MODE MANUAL
INPut:INHibit?
Function
Queries all settings related to the inhibit
function.
Syntax
:INPut:INHibit?
Example
:INPUT:INHIBIT?
→
:INPUT:INHIBIT:LEVEL ZERO
INPut:INHibit:LEVel
Function
Sets the inhibit level or queries the current
setting.
Syntax
:INPut:INHibit:LEVel {ZERO|TTL|TTL10}
:INPut:INHibit:LEVel?
Example
:INPUT:INHIBIT:LEVEL ZERO
:INPUT:INHIBIT:LEVEL?
→
:INPUT:INHIBIT:LEVEL ZERO
3.9 INPut Group