YOKOGAWA SMARTDAC+ GM Скачать руководство пользователя страница 16

1-11

TI 77C01H01-01EN

Oct. 17, 2018-00

P1

P2

P3

1S

1L

3S

3L

(1)(2)(3)

(1)(2)(3)

(4)

3 phase 3 wire

R  

S

T

FUSE

FUSE

Power supply

Load

k

K

K

l

L

L

k

l

(1) CT orientation

×  

 

Power monitor terminal

Current input

Voltage input

Check the installation for other phase and wiring schemes as well.

  Wiring RS-485 communication

Wiring for communication between the Data Acquisition System GM and the UPM100 Power 

Monitor (RS-485 4 wire <=> 2 wire)

(FG)

Approximately 1.2 km maximum length (up to 31 units)

Ethernet

RS-485

S GM

(RDA-) (RDB+) (SDA-) (SDB+) (SG)

(A-) (B+) (SG)

(A-) (B+) (SG)

(A-) (B+) (SG)

RS-485

RS-485

UPM100

GM10

UPM100

UPM100

Terminator

120Ω

Terminator

120Ω

PC

Содержание SMARTDAC+ GM

Страница 1: ...MARTDAC GM and UPM100 Power Monitoring Package Wiring and Setup Guide TI 77C01H01 01EN TI 77C01H01 01EN Copyright Oct 2018 1st Edition Oct 2018 Yokogawa Electric Corporation 2 9 32 Nakacho Musashino s...

Страница 2: ...he contents of this manual in part or in whole be transcribed or copied without our permission Trademarks Our product names or brand names mentioned in this manual are the trademarks or registered tra...

Страница 3: ...Blank Page...

Страница 4: ...Guide CONTENTS TI 77C01H01 01EN TOFC How to use this manual i 1 Overview 1 1 2 Device configuration and related documents 1 2 3 Package specifications 1 3 4 UPM connection settings 1 4 5 Universal Vie...

Страница 5: ...Blank Page...

Страница 6: ...g system designed for smallscale applications It consists of the SMARTDAC GM the UPM100 Power Monitor and a CTW fixed wire type current transformer You can also select instruments and communication fu...

Страница 7: ...3 01E Models CTW10 CTW20 CTW100 and CTW130 Clamp on Current Transformers GS 77C01W02 01E User s manual Title Number Data Acquisition System GM User s Manual IM 04L55B01 01EN Data Acquisition System GM...

Страница 8: ...or Ethernet RS 485 Ethernet converter when using VJET Functions You can acquire watt hours from the power monitor UPM100 and display digital values intetration bar graphs and integral trends on the we...

Страница 9: ...f integral graphs You can use it with the hardware configurator version 4 05 01 or later and the web application version 4 05 01 or later Hardware Configurator Web Application UPM Connection settings...

Страница 10: ...tarts From this screen you can load UPM connection settings files Apply settings on the GM Settings are complete Completed settings can be saved to a UPM Connection Settings file Setup complete Start...

Страница 11: ...ernet GM UPM 1 Unit 2 Address 2 UPM 2 Unit 3 Address 3 UPM 3 Unit 4 Address 4 IP 192 168 1 99 IP 192 168 1 100 VJET Server gateway Client IP 192 168 1 1 Unit 1 Address 1 RS 485 and Ethernet connection...

Страница 12: ...he date of the integral graph to display Legend and cursor value display area The legend and cursor value are docked on the right edge of the screen Two graph comparison You can compare two integral g...

Страница 13: ...lines P1 and P3 n Setting the VT and CT ratios What are the VT and CT ratios Generally when measuring high voltage and high current transformer signals you use a VT voltage transformer and CT current...

Страница 14: ...ance Connected device Device input loss Wiring loss Checking the wiring distance and material The wiring loss increases in proportion with the wiring distance The total wiring length is the round trip...

Страница 15: ...secondary side connection short The CTW series of fixed wire type current transformers have built in hazardous voltage suppressors that function when the secondary side is open In the unlikely case t...

Страница 16: ...Check the installation for other phase and wiring schemes as well n Wiring RS 485 communication Wiring for communication between the Data Acquisition System GM and the UPM100 Power Monitor RS 485 4 wi...

Страница 17: ...of quantities of vapor gas and other substances You can add a pulse module for acquiring flow meter pulses and measure flow quantities of power vapor and gas System architecture SMARTDAC GM UPM100 Fl...

Страница 18: ...AC GM and UPM100 Power Monitoring Package Wiring and Setup Guide Manual number TI 77C01H01 01EN Oct 2018 1st Edition Newly published Written by Yokogawa Electric Corporation Published by Yokogawa Elec...

Страница 19: ...Blank Page...