![YOKOGAWA ScopeCorder DL350 Скачать руководство пользователя страница 118](http://html.mh-extra.com/html/yokogawa/scopecorder-dl350/scopecorder-dl350_user-manual_911346118.webp)
6-11
IM DL350-01EN
Reference Times (Time, T1, T2)
You can set the reference times (Time, T1, and T2) within the following ranges.
Setting
Selectable Range
Resolution
Time
0.02 μs to 10000000.00 μs (10 s)
0.01 μs
T1
0.01 μs to 9999999.99 μs
T2
0.02 μs to 10000000.00 μs (10 s)
Triggering may not function properly when the interval between achievement times is less than 0.01 μs or
when the duration of the achievement time is less than 0.01 μs (Typical).
Trigger Position (Position), Trigger Delay (Delay)
, Trigger delay
Wave Window Trigger (Enhanced)
This is available only in scope mode.
The instrument creates real-time templates (Wave Window) using a number of cycles directly preceding the
current waveforms. The instrument compares the current waveforms to the real-time templates and triggers if
one of the current waveforms falls outside of its real-time template.
Reference cycles (example: the waveforms
for the four previous cycles)
Current waveform
Real-time template
Average waveform
±
tolerance
Trigger
Average of 1, 2, or 4
cycles
The instrument compares the
current waveform to the
real-time template and triggers
if the current waveform falls
outside of the real-time
template.
Measured waveform
Tolerance
Tolerance
Trigger Source
You can select the trigger sources from CH1 to CH6. The modules that you can use as sources for the Wave
Window trigger are listed below. You cannot use other modules or temperature-measurement channels as
sources.
720211 (HS100M12)
720250 (HS10M12)
Sub channel 1 of 720254 (4CH 1M16)
720268 (HV (with AAF, RMS))
701261 (UNIVERSAL) (only voltage
measurement)
701262 (UNIVERSAL (AAF)) (only voltage
measurement)
701270 (STRAIN_NDIS)
701271 (STRAIN_DSUB)
701275 (ACCL/VOLT)
Template Channels (Condition)
Select which trigger sources to use to make real-time templates. The instrument triggers if the condition of even
one of the channels is met.
• ON: Use
• OFF: Don’t use
6 Triggering