3. INSTALLATION

3.1 Typical installation

The SC25V sensor can be installed in-line, in a 

bypass loop or in an immersion assembly. For

best results the SC25V should be mounted 

with the process fluid flowing towards the 

sensor and positioned at least 15° above the 

horizontal plane to eliminate air bubbles in the 

glass bulb. See fig. 6 section 3.3

3.2 Preparing the sensor for use

Remove the sensor from its shipping box and 

slide off the “wet pocket” (the flask including 

the rubber ring). This wet pocket is filled with 

a weak acid and saline solution to prevent the 

sensor from drying out during storage and 

making sure it is ready for immediate use.

During shipment, electrolyte in the sensor 

could be dislocated. To correct this, place the 

sensor upright for 24 hours.

Before mounting the sensor, it should be 

validated with buffer solutions and calibrated 

when necessary. The specific calibration 

procedure is described in the Instruction 

Manual of the pH transmitter, but a general 

procedure is also provided in section 5 of this 

manual.

3.3 Mounting the sensor

The SC25V can be mounted using: 

• PR10 retractable fitting (see fig 2).

•  FF20 flow fitting / FS20 subassembly / FD20 

immersion fitting (see fig 4)

•  For mounting in Yokogawa fittings, a PG13.5 

to M25 adapter is available in different 

materials (see fig 3). More information in 

chapter 7.

•  PD20 immersion fitting / PF20 flow fitting / 

PS20 flow fitting (see fig 5)

•  FF40 fitting with PG13.5 adapter (K1523JA/

JC) or the small flow fitting K1598AC with 

PG13.5 adapter (K1523JB/JD), (see fig 7).

PR10

O-RING VITON 11x3

SC25V

O-RING 11 x 3

SLIDE RING 

RYTON

SC25V

K1520JN, K1520JP

K1500DV

FD20 / FF20 / FS20

2 x O-RING  11x3

K1520JN

K1520JP

K1500DV

SC25V

PD20 / PF20 / PS20

O-RING 11 x 3

SLIDE RING 

RYTON

SC25V

Содержание SC25V-AGP25-120

Страница 1: ...IM12B06J01 40E E 4th edition Instruction Manual YOKOGAWA Model SC25V Combined 12mm sensor pH Ref LE and Temperature...

Страница 2: ...2 IM12B06J01 40E E...

Страница 3: ...2 5 Dynamic specifications 5 2 6 Sensor Dimensions 6 2 7 Equipment and protective systems for use in potentially explosive atmospheres 6 3 INSTALLATION 7 3 1 Typical installation 7 3 2 Preparing the...

Страница 4: ...typical warranty period and the conditions of sales relating to the original purchase order should be consulted Damage caused by wear and tear inadequate maintenance corrosion or by the effects of ch...

Страница 5: ...emperature control 2 4 Operating range pH 0 to 14 Temperature Type G glass 10 to 80 C 14 to 176 F Type L glass 15 to 130 C 59 to 266 F Pressure 0 to 10 bar 0 to 145 psi Conductivity 10 S cm Storage te...

Страница 6: ...hout ID chip the I O signals are from to an associated intrinsically safe certified pH transmitter e g Yokogawa transmitter Model FLXA21 series or Yokogawa transmitter Model PH202S series For the mode...

Страница 7: ...he sensor upright for 24 hours Before mounting the sensor it should be validated with buffer solutions and calibrated when necessary The specific calibration procedure is described in the Instruction...

Страница 8: ...1523JB JD PG13 5 adapter K1523JA JC Conductivity Flow Fitting FF40 SC4A option FF K1523JA K1598AC Flow fitting FF40 with K1523JA JC Adapter to fit sensors with a PG13 5 process connection in FF40 FS40...

Страница 9: ...le etc 5 Rinse sensor thoroughly with clean water Carefully slap out residual fluid drops 6 Immerse the sensor in the second buffer pH 4 01 buffer or pH 9 18 buffer recommended 7 Repeat Step 3 Most in...

Страница 10: ...O RINGS EPDM 11X3 6 PCS K1500BZ O RINGS VITON 11X3 6 PCS K1500GR O RINGS SILICON 11X3 8PCS K1524AA Set of O ring 11x3 and slide ring Ryton FP20 R12 Blind plug set for 1 hole 7 2 Adapters Prod No Desc...

Страница 11: ...o o o o o o o x Ammonia in water 30 x x x o o o x x o o o o o o o o x Ammonium chloride sat o o o o o o o o o o o o o o o o o o Zinc chloride 50 o o o o o o o o o o o o o o o o o o Iron III chloride 5...

Страница 12: ...herlands www yokogawa com eu YOKOGAWA ELECTRIC CORPORATION World Headquarters 9 32 Nakacho 2 chome Musashino shi Tokyo 180 8750 Japan www yokogawa com YOKOGAWA CORPORATION OF AMERICA 2 Dart Road Newna...

Отзывы: