IM 12D7B3-E-H
Parameter setting 5-17
Code
Display
Function
Function detail
X
Y
Z
Default values
User interface
50
*RET
Auto return
Auto return to measuring mode Off
0
Auto return to measuring mode On
1
1
On
51
Not used
52
*PASS
Passcode
Maintenance passcode Off
0
0.0.0
Off
Note # = 0 - 9, where
Maintenance passcode On
#
Commissioning passcode Off
0
Off
1=111, 2=333, 3=777
Commissioning passcode On
#
4=888, 5=123, 6=957
Service passcode Off
0
Off
7=331, 8=546, 9=847
Service passcode On
#
53
*Err.01
Error setting
Polarization too high
Soft/Hard
0/1
1
Hard
*Err.05
Shorted measurement
Soft/Hard
0/1
1
Hard
*Err.06
Open measurement
Soft/Hard
0/1
1
Hard
*Err.07
Temperature sensor open
Soft/Hard
0/1
1
Hard
*Err.08
Temp. sensor shorted
Soft/Hard
0/1
1
Hard
*Err.13.
USP limit exceeded
Soft/Hard
0/1
0
Soft
54
*E5.LIM
E5 limit setting
Maximum conductivity value
250
mS
(Minimum resistivity value)
0.004 k
Ω
*E6.LIM
E6 limit setting
Minimum conductivity value
1.000 µS
(Maximum resistivity value)
1.000 M
Ω
55
*%
Display mA in w/w%
mA-range displayed in w/w% off
0
Off
mA-range displayed in w/w% on
1
*0%
Set 0% output value in w/w%
*100%
Set 100% output value in w/w%
56
*DISP
Display resolution
Auto ranging display
0
0
Auto
Display fixed to X.XXX µS/cm or M
Ω
.cm
1
Display fixed to XX.XX µS/cm or M
Ω
.cm
2
(2)
Display fixed to XXX.X µS/cm or M
Ω
.cm
3
Display fixed to X.XXX mS/cm or k
Ω
.cm
4
Display fixed to XX.XX mS/cm or k
Ω
.cm
5
Display fixed to XXX.X mS/cm or k
Ω
.cm
6
Display fixed to XXXX mS/cm or k
Ω
.cm
7
57
*USP
USP setting
Disable the E13 (USP limit exceeded)
0
0
Off
Enable the E13 (USP limit exceeded)
1
58-59
Not used
Содержание SC202G(S)
Страница 2: ......
Страница 24: ...IM 12D7B3 E H ...
Страница 48: ...IM 12D7B3 E H ...
Страница 52: ...IM 12D7B3 E H ...
Страница 54: ...IM 12D7B3 E H ...
Страница 60: ...IM 12D7B3 E H ...
Страница 71: ...Test Certificate 12 3 IM 12D7B3 E H ...