IM 12D7B2-E-H

31

7-6. Output table for non-linear range

ACCESS-CODE : 04 (see §7-2)
DISPLAY

:  0% to 100%

Adjustment

: A 21 step table can be

programmed from this
entry ONLY IF it was set
in the previous CODE 03.
It is possible to set up a
bi-linear, semi-logarith-
mic or hyperbolic scale
to cope with retro fitting
on existing installations.
Also linear output to con-
centration related curves
can be programmed.

Explanation:

First you will be presented with the question
YES/NO to give the opportunity to skip valu-
es from the table. To get a 10 step table you
would skip every other value and only ans-
wer YES to enter a value. The instrument
uses interpolation between entered values.
Press NO to skip values. Press YES to
adjust value. Now adjust the value by pres-
sing > and ^.
The units can be adjusted as well as the
decimal point position.
Examples are mentioned in §7-17.
Other possibilities: programming ”% by
weight” indication on second display line:
refer to §7-13.

Percentage of mA-output range versus
percentage of conductivity range.

Examples:
A = bilinear
B = hyperbolic (2 decades)
C = logarithmic (2 decades)
D =linear

NOTE:
Curves are not selectable, but can be pro-
grammed from the tables at right.

Example of output tables

bi-lin

= bi-lineair over 2 decades

log 2

= logarithmic over 2 decades

log 3

= logarithmic over 3 decades

hyp 2

= hyperbolic over 2 decades

hyp 3

= hyperbolic over 3 decades

NOTE:
Multiply the values from the table with
appropriate factors to get the end-scale
value you want.

Special concentration tables can be pro-
grammed by the user.

mA-output range

% of conductivity range

Output

bi-lin

log 2

log 3

hyp 2

hyp 3

0%

0.0

1.0

0.10

1.00

0.10

5%

1.0

1.3

0.14

1.20

0.27

10%

2.0

1.6

0.20

1.82

0.43

15%

3.0

2.0

0.28

1.90

0.61

20%

4.0

2.5

0.40

2.00

0.83

25%

5.0

3.2

0.56

3.75

1.10

30%

6.0

4.0

0.79

4.80

1.36

35%

7.0

5.0

1.12

5.92

1.68

40%

8.0

6.3

1.58

7.00

2.05

45%

9.0

7.9

2.24

8.31

2.49

50%

10.0

10.0

3.16

10.00

3.00

55%

20.0

12.6

4.47

11.85

3.66

60%

30.0

15.8

6.31

14.00

4.33

65%

40.0

20.0

8.91

16.65

5.22

70%

50.0

25.1

12.6

19.50

6.80

75%

60.0

31.6

17.8

23.80

8.25

80%

70.0

39.8

25.1

29.55

11.0

85%

80.0

50.1

35.5

36.70

14.8

90%

90.0

63.1

50.1

48.50

21.8

95%

100.0

79.4

70.8

68.60

36.5

100%

110.0

100.0

100.0

100.00

100.0

Содержание SC200

Страница 1: ...Instruction Manual Model SC200 2 wire Conductivity or Resistivity transmitter IM 12D7B2 E H 4th edition...

Страница 2: ...E VALUE DISPLAY MESSAGE DISPLAY KEY PROMPT FLAGS SELECTION KEYS YES Accept setting NO change to new setting L L SELECT MODE MEASURE MAINTENANCE Can be used to escape program at any time SELECT MODE ME...

Страница 3: ...11 4 2 Output range adjustment OUTPUT 12 4 3 Set up Hold function SET HOLD 14 4 4 Temperature compensation TEMPERATURE 16 4 5 Sensor selection and diagnostics 18 5 MAINTENANCE 20 5 1 Calibration CAL 2...

Страница 4: ...t 38 9 2 Resistivity measurement 38 9 3 Maintenance of the transmitter 38 10 SOFTWARE VERSION 3 0 COMPARED TO VERSIONS 1 AND 2 39 10 1 Cell constant 39 10 2 Matrix compensation 39 10 3 Matrix temperat...

Страница 5: ...functions In general a conductivity loop can be set up for different purposes To be part of a total process control sys tem To indicate dangerous limits of a process To monitor product quality 1 2 Re...

Страница 6: ...tion EMI Class B RFI less than 2 at 5 V m for 20 MHz to 1 GHz G Display method Custom liquid crystal display Main display 31 2digit 12 5 mm high Message display 6 alphanumeric characters 7 mm Special...

Страница 7: ...h the operator by giving messages on the second line of the display area and indicating which keys are to be pressed in the display too The keys are scanned continuously and the actions are taken imme...

Страница 8: ...ot mounted in direct sunlight and severe weather conditions maintenance activities are possible no corrosive atmospheres The ambient temperature and humidity should be the limits of the specifications...

Страница 9: ...IM 12D7B2 E H 5 Universal pipe wall mounting Wall mounting Pipe mounting vertical Pipe mounting horizontal 2 I D pipe 2 I D pipe 115 200 70 4x 10 2x 6 6 56 77...

Страница 10: ...e is not to be mounted in tracks together with high voltage and or power switching cables use only the standard conductivity cable with a maximum length of 10 metre mount the transmitter within the di...

Страница 11: ...SENSOR TEMPERATURE SENSOR CEL C 10 cm C 1 cm C 0 1 cm C 0 01 cm CEL e g Use cable WU40 LH Max recommended length 10 m Use Cells SC41 or SC42 Range selection Use of 4 electrode cell recommended e g Us...

Страница 12: ...on a grounded surface e g a metal frame fixed in the soil the shield of the 2 wire cable may NOT be connected to ground at the distributor If the transmitter is mounted on a non conducting surface e g...

Страница 13: ...OTE The outside earth terminal should be connected to site ground by a large area conductor e g a flat earth strip for best protection against interference Sensor Distributor Output Supply Dependance...

Страница 14: ...x 0 66 W Zener barrier with resistor Vmax 28 V lmax 93 3 mA Pmax 0 66 W Protective earth Safe area Protective earth Hazardous area Supply Zone 0 of 1 Zone 1 Eex ib Certified distributor with input out...

Страница 15: ...hold function on off 5 1 5 2 5 3 Adjusting the output range Activating the hold function Temperature compensation 4 2 4 3 4 4 Fine tuning the performance 7 OUTPUT SET HOLD TEMPERATURE SERVICE COMMISSI...

Страница 16: ...span Select digit to adjust or decimal point and units Adjust digit Confirm adjusted value ENT L Return to measurement select output function Select OUTPUT YES NOTE If the output is commissioned to be...

Страница 17: ...ity at the end of the output range which corresponds to 20 mA When programming a range which does not start at zero conductivity take care to respect the minimum value corresponding to 60 of the maxim...

Страница 18: ...HOLD Access commissioning menu SELECT choice Select SET HOLD Move pointer to SET HOLD NO YES Activate de activate HOLD Change setting Conform setting NO YES Select HOLD fixed or last Change selection...

Страница 19: ...p the output frozen during the following events Access to the menu Access to the calibration mode Switching it from the MODE menu HOLD is signalled in the display by a special field The operator is pr...

Страница 20: ...HOLD TEMPERATURE SERVICE YES NO YES NO MODE OUTPUT SET HOLD TEMPERATURE SERVICE YES NO YES NO MODE O U T P U T S E T H O L D T E M P S E R V I C E EN T ENT TEMP MAN mS cm Access the maintenance menu S...

Страница 21: ...l NaCl solutions In the high conductivity region the effect of temperature decreases and the compensation factor is decreased too In these extreme regions the tempera ture compensation factor is calcu...

Страница 22: ...steel epoxy resin PTFE and PVDF The cell determines the principle used to measure the conductivity there is a distinc tion between 2 electrodes and 4 electrodes cells The selection of 2 or 4 electrod...

Страница 23: ...IM 12D7B2 E H 19...

Страница 24: ...ter to display NO YES End calibration End calibration and return to measurement Continue routine to calibrate second point repeat step 2 YES NO MODE Conform start of adjustment YES Select digit to adj...

Страница 25: ...mperature is stabi lised to the reference temperature of the in strument usually 25 C The conductivity of the solution is taken from tables Alternatively the instrument may be calibra ted in an unspec...

Страница 26: ...YES NO YES NO MODE MODE M E A S U R E C A L D I S P L AY H O L D YES NO YES NO MODE MODE Access maintenance mode Select Display routine Select display Move pointer to DISPLAY NO YES Read data To read...

Страница 27: ...temperature coefficient is program med The choice of temperature units is done from the Service level Error messages have priority over other messages Note Percent by weight is only visible when activ...

Страница 28: ...Display will blink HOLD and YES NO 1 Access hold 2 Switch hold on off M E A S U R E C A L D I S P L AY H O L D YES NO YES NO MODE MODE M E A S U R E C A L D I S P L AY H O L D YES NO YES NO MODE MODE...

Страница 29: ...s switched on when you press YES when HOLD and YES NO are blinking When you press NO hold will be switched off A flag is kept in memory and an indication is made in the upper left corner of the displa...

Страница 30: ...reduced by increasing the measu ring frequency At all values the instrument checks the sig nal from the cell to search for distortion which is typical of capacitive or polarisation errors If there is...

Страница 31: ...ure sensor selected Program sensor type Fault in connections Check connections E8 Temperature too low Wrong temperature sensor selected Program sensor type Fault in connections Check connections E10 E...

Страница 32: ...ogram ming again 28 Code Routine Use Chapter 01 Temperature sensors and units Select sensor type 7 3 Select C or F 02 Reference temperature Adjust value 7 4 03 Output table or linear Select which to u...

Страница 33: ...ription 1 Access service 2 Enter code to select required function 3 Adjust setting see individual routines OUTPUT SET HOLD TEMPERATURE SERVICE YES NO YES NO MODE O U T P U T S E T H O L D T E M P S E...

Страница 34: ...set the correct number for it Default value is Pt 1000 for Yokogawa sen sors type SC42 Default C Explanation All measured values are compensated for temperature to refer to comparable values at 25 C...

Страница 35: ...e of conductivity range Examples A bilinear B hyperbolic 2 decades C logarithmic 2 decades D linear NOTE Curves are not selectable but can be pro grammed from the tables at right Example of output tab...

Страница 36: ...0 1 cm l 7 8 Auto return function ACCESS CODE 06 see 7 2 DISPLAY RET Adjustment X Explanation As a safeguard against long maintenance jobs or inadvertently pushing a button it is possible to let the s...

Страница 37: ...sary This is possi ble by sending a special signal over the 2 wire connection to the receiving instrument A signal of 22 mA is used because it is out side the normal analogue range of 4 to 20 mA If at...

Страница 38: ...uses interpolation between the entered values As the relation between percent by weight concentration and con ductivity is not linear in most cases the pos sibility to linearize the output to concent...

Страница 39: ...scode is selected for an operation level access to that level can only be obtained after ente ring the passcode The display will show a message PASS to indicate the entry of the passcode NOTE For the...

Страница 40: ...n body to EEx ia II C or EEx ib ii C where the output current is limited by a resistor R directly in the output line such as that Imax out Vmax out R and not exceeding the following char acteristics V...

Страница 41: ...requirements Maximum safe area voltage should not exceed 250 VRMS NON HAZARDOUS LOCATION Class I Division I Groups ABCD Vmax 31 5 V Imax 93 3 mA Ci 5 nF Li 16 5 uH ControlDrawing for EXA 200 instrumen...

Страница 42: ...transmitter can be reassembled now following the opposite sequence as described above 9 2 Resistivity measurement As a resistivity measuring transmitter the EXA SC200 has the same function as describ...

Страница 43: ...equently there is no need for a 100 coverage If 9 is selected the temperature compensa tion range for the user adjustable matrix can be set in access code 15 In access codes 16 and 19 the specific con...

Страница 44: ...nge is 50 C Error code 17 indicates a temperature range of less than 50 C 10 4 Matrix data ACCESS CODE 15 16 17 18 19 Display L1 T1 Adjustment After adjusting the range in access code 14 it is necessa...

Страница 45: ...6 the error codes marked with an asterisk normally will trig ger a number of actions If soft alarm action is selected the FAIL message flag is flashing The output func tions will not go into BURN OUT...

Страница 46: ...6 1 7 3 3 100 0 7681 0 98 1 3 2 1 3 9 Ammonia 2 0 0 00975 0 122 0 23 0 43 0 82 Conductivity 25 0 0559 0 25 0 50 1 0 2 0 S cm 50 0 1752 0 41 0 80 1 62 3 35 75 0 4092 0 65 1 13 2 3 4 8 100 0 7681 1 1 52...

Страница 47: ...IM 12D7B2 E H...

Страница 48: ...22 BELGIUM Yokogawa Belgium N V S A Minervastraat 16 1930 ZAVENTEM Tel 32 2 719 55 11 Fax 32 2 725 34 99 FRANCE Yokogawa Contr le Bailey S A V lizy Valley 18 20 Rue Grange Dame Rose 78140 VELIZY VILLA...

Отзывы: