General Instruction Manual
Specifications
Ambient conditions
88 / 90
IM 01U10B00-00EN-R, 3rd edition, 2018-07-09
13 Specifications
Specifications for sensor and transmitter are listed in the General Specifications of the
corresponding product family.
13.1 Ambient conditions
Rotamass Total Insight can be used at demanding ambient conditions.
In doing so, the following specifications must be taken into account:
As ambient temperature is intend the air surrounding the device.
Allowed ambient and storage temperature of Rotamass Total Insight depends on the be-
low components and their own temperature limits:
▪ Sensor
▪ Transmitter
▪ Connecting cable between sensor and transmitter (for remote design type)
Ambient
temperature
Maximum ambient temperature range
1)
integral type
:
-40 – 60 °C (-40 – 140 °F)
remote type
with standard cable
(option L
␣␣␣
):
Sensor
2)
:
-50 – 80 °C (-58 – 176 °F)
Transmitter: -40 – 60 °C (-40 – 140 °F)
with fire retardant cable
3)
(option Y
␣␣␣
):
Sensor
2)
:
-35 – 80 °C (-31 – 176 °F)
Transmitter: -35 – 60 °C (-31 – 140 °F)
1)
If the device is operating outdoors make sure that the solar irradiation does not increase
the surface temperature of the transmitter higher than the allowed maximum ambient
temperature. Transmitter display has limited legibility below -20 °C (-4 °F)
2)
Check derating for high fluid temperature, see Process fluid temperature range,
Process conditions and Allowed ambient temperatures for sensor.
3)
Lower temperature specification valid for fixed installation only
Storage
temperature
Maximum storage temperature range
integral type
-40 – 60 °C (-40 – 140 °F)
remote type
with standard cable
(option L
␣␣␣
):
Sensor:
-50 – 80 °C (-58 – 176 °F)
Transmitter: -40 – 60 °C (-40 – 140 °F)
with fire retardant cable
(option Y
␣␣␣
):
Sensor:
-35 – 80 °C (-31 – 176 °F)
Transmitter: -35 – 60 °C (-31 – 140 °F)
Further
ambient conditions
Ranges and specifications
Relative humidity
0 – 95 %
IP code
IP66/67 for transmitters and sensors when
using the appropriate cable glands
Allowable pollution degree in surrounding
area acc. EN 61010-1
4 (in operation)
Vibration resistance acc. IEC 60068-2-6
Transmitter: 10 – 500 Hz, 1g
Sensor: 25– 100 Hz, 4g