Troubleshooting
If it appears even though a sensor with a built-in temperature element is used, the
temperature measuring circuit may have broken. In this case the meter performs
temperature compensation assuming that the solution temperature is 25°C.
Therefore, the bigger the difference between the actual temperature of the solution
and 25°C, the bigger the error between the displayed measured value and the true
value. For reliable measurement replace the sensor with a good one.
This mark will also appear if the sensor is not connected properly. Make sure that the
connectors are securely connected.