IM 12B03D02-01E
4-1
4. Calibration
4. Calibration
Calibration using standard solutions means to measure the pH value of a certified
standard solution and to adjust the pH meter so it reads the same value as the certified
value of the standard solution. The PH72 meter can be calibrated automatically or
manually.
The PH72 meter must be calibrated before measurement if:
• it has been stored for a long period;
• the electrode has been cleaned; or
• otherwise necessary.
Precautions
(1) Use certified standard solutions. Using deteriorated standard solutions will result
in inaccurate calibration. Standard solutions are available as spare parts (see
Section 1.8).
Pour a portion of the standard solution into a calibration bottle (supplied) and use
for calibration. Do not reuse the portion. Discard it and use a fresh portion for
another calibration.
(2) Do not press
CAL
key unless calibration is needed. Otherwise, saved calibration
results may be changed.
Before Calibration
The following should be checked and set before calibration.
(1) Contamination of the sensor
Make sure that no dirt or deposits are present on the sensor.
(2) Temperature setting
When using a sensor without a built-in temperature element, i.e., a needle type or test
tube size pH sensor, the temperature of a standard solution to be used should be set into
the meter. (Refer to Section 5.3 (2), “Manual temperature setting (M.tP) panel.”)
(3) Battery condition
Check the battery condition indicator for remaining life. If the indicator is flashing,
calibration cannot be performed. Replace the batteries. (Refer to Section 2.1, “Installing
the Batteries.”)
Содержание PH72-21-E-AA
Страница 1: ...User s Manual IM 12B03D02 01E Model PH72 Personal pH ORP Meter IM 12B03D02 01E 3rd Edition...
Страница 7: ...IM 12B03D02 01E 6 Contents...
Страница 23: ...IM 12B03D02 01E 2 6 2 Preparation...
Страница 35: ...IM 12B03D02 01E 4 8 4 Calibration...
Страница 49: ...IM 12B03D02 01E 5 14 5 Keypad and Display Functions...
Страница 57: ...IM 12B03D02 01E 6 8 6 Maintenance...
Страница 85: ...IM 12B03D02 01E 6 Appendix...
Страница 87: ...IM 12B03D02 01E MSDS 2 MSDS...
Страница 88: ...IM 12B03D02 01E MSDS 3 MSDS...
Страница 89: ...IM 12B03D02 01E MSDS 4 MSDS...
Страница 90: ...IM 12B03D02 01E MSDS 5 MSDS...
Страница 91: ...IM 12B03D02 01E MSDS 6 MSDS...
Страница 92: ...IM 12B03D02 01E MSDS 7 MSDS...
Страница 93: ...IM 12B03D02 01E MSDS 8 MSDS...
Страница 94: ...IM 12B03D02 01E MSDS 9 MSDS...
Страница 95: ...IM 12B03D02 01E MSDS 10 MSDS...
Страница 96: ...IM 12B03D02 01E MSDS 11 MSDS...
Страница 97: ...IM 12B03D02 01E MSDS 12 MSDS...
Страница 98: ...IM 12B03D02 01E MSDS 13 MSDS...
Страница 99: ...IM 12B03D02 01E MSDS 14 MSDS...
Страница 100: ...IM 12B03D02 01E MSDS 15 MSDS...
Страница 101: ...IM 12B03D02 01E MSDS 16 MSDS...
Страница 102: ...IM 12B03D02 01E MSDS 17 MSDS...
Страница 103: ...IM 12B03D02 01E MSDS 18 MSDS...
Страница 104: ...IM 12B03D02 01E MSDS 19 MSDS...
Страница 105: ...IM 12B03D02 01E MSDS 20 MSDS...
Страница 106: ...IM 12B03D02 01E MSDS 21 MSDS...
Страница 107: ...IM 12B03D02 01E MSDS 22 MSDS...