IM 12B03D01-01E
6-5
6. Maintenance
6.6 Cleaning and Drying Connectors
Deteriorated insulation between connector pins can cause inaccurate readings. To remove
stains and/or moisture that may cause deteriorated insulation, clean the connector with a
dry cloth or a cloth moistened with anhydrous alcohol. If necessary, use a dryer.
Wipe off stains and/or moisture on
meter connector with a dry cloth.
Check that there are no stains on
the O-ring.
O-ring
Hot air
Use a dryer if necessary to
remove moisture from
connector of sensor cable.
F060601.eps
CAUTION
Use anhydrous alcohol to clean the connectors so no moisture remains. Dry connectors
completely.
Содержание PH71
Страница 1: ...User s Manual IM 12B03D01 01E Model PH71 Personal pH Meter IM 12B03D01 01E 3rd Edition ...
Страница 7: ...IM 12B03D01 01E 6 Contents ...
Страница 17: ...IM 12B03D01 01E 1 10 1 Outline ...
Страница 23: ...IM 12B03D01 01E 2 6 2 Preparation ...
Страница 35: ...IM 12B03D01 01E 4 8 4 Calibration ...
Страница 49: ...IM 12B03D01 01E 5 14 5 Keypad and Display Functions ...
Страница 57: ...IM 12B03D01 01E 6 8 6 Maintenance ...
Страница 73: ...IM 12B03D01 01E 8 10 8 Technical Information ...
Страница 79: ...IM 12B03D01 01E 6 Appendix ...
Страница 81: ...IM 12B03D01 01E MSDS 2 MSDS ...
Страница 82: ...IM 12B03D01 01E MSDS 3 MSDS ...
Страница 83: ...IM 12B03D01 01E MSDS 4 MSDS ...
Страница 84: ...IM 12B03D01 01E MSDS 5 MSDS ...
Страница 85: ...IM 12B03D01 01E MSDS 6 MSDS ...
Страница 86: ...IM 12B03D01 01E MSDS 7 MSDS ...
Страница 87: ...IM 12B03D01 01E MSDS 8 MSDS ...
Страница 88: ...IM 12B03D01 01E MSDS 9 MSDS ...
Страница 89: ...IM 12B03D01 01E MSDS 10 MSDS ...
Страница 90: ...IM 12B03D01 01E MSDS 11 MSDS ...
Страница 91: ...IM 12B03D01 01E MSDS 12 MSDS ...
Страница 92: ...IM 12B03D01 01E MSDS 13 MSDS ...
Страница 93: ...IM 12B03D01 01E MSDS 14 MSDS ...
Страница 94: ...IM 12B03D01 01E MSDS 15 MSDS ...
Страница 95: ...IM 12B03D01 01E MSDS 16 MSDS ...