Sometimes you will want to know how you sensor is 
functioning. To view the sensors zero mV output 
from the electrode (AS), Slope (SL) and 

Impedance 1 and Impedance 2. From the Main 
screen press the I icon in the circle. 
 
 

 
 
 
 
 

The following screen will pop up press the 
diamond next to NEXT.  
 
 

 
 
 
 
 

 
 
 

Then… the data screen will appear. The zero is 
the AS, the slope is the SL, and the impedance 

2 is the impedance of the reference junction. 
The sensor mV reading is a continuous reading 
of what the sensor is seeing in normal 
measureing mode  
 

 
 
To return to normal measureing mode press the 
house icon. 
 

 
 
 
 
 
 

 

Содержание PH450

Страница 1: ...traceable buffers are used and the default values in the preprogrammed Buffer Tables have not been changed an error of 14 pH 7 00 6 86 14 will occur To eliminate this situation either pure NBS buffers should be used or the Buffer Tables in the instrument should be reprogrammed to match the values of the buffers being used For further information regarding NBS vs NBS Traceable buffers refer to the ...

Страница 2: ...ration Wash select it 3 Next selected the diamond next to Automatic zer slope 4 On the following screen in the box at the bottom use the scroll key to select the 6 9 buffer 5 The main display will start flashing and continues to flash until the reading has stabilized for 5 seconds and does not change by more than 02 pH units ...

Страница 3: ...7 00 buffer and rinse it thoroughly in deionized water then blot the electrodes to remove any excess water Immerse the electrode assembly in a sample cup of 4 buffer solution 8 On the following screen in the box at the bottom use the scroll key to select the 4 0 buffer 9 The main display will start flashing and continues to flash until the reading has stabilized for 5 seconds and does not change b...

Страница 4: ...ore it in memory Press the diamond next to Calibration complete 11 To have the analyzer finish and accept calibration data press the diamond next to Accept Data 12 The analyzer will automatically return you to the original Calibration screen To return to main operating mode press the house icon to return to the main screen ...

Страница 5: ... the Main screen press the I icon in the circle The following screen will pop up press the diamond next to NEXT Then the data screen will appear The zero is the AS the slope is the SL and the impedance 2 is the impedance of the reference junction The sensor mV reading is a continuous reading of what the sensor is seeing in normal measureing mode To return to normal measureing mode press the house ...

Отзывы: