<8. Parameter Settings>
8-1
IM 11M10B01-01E
8. Parameter Settings
Table
8.1
Parameter
Setting
List
(Customer
Settings)
Menu
Parameter 1
Parameter 2
Parameter 3
Function
detail
Defaults
Description
Customer
Settings
Reference
section
Output
range
(4-20
mA
DC)
“rnG”
“Auto”
Auto range
“1”
(Note
1)
6.1.1
“MAn”
Manual range
0-1000 ppm
“FrEE”
Partial range
0-1000 ppm
“Fr
.Lo”
Low limit
0 ppm
“Fr
.HI”
High limit
1000 ppm
Secondary
output
“oUt2”
“1”
/
“5”
/
“10”
Voltage output
“1”
0-1V DC
6.1.2
Hold
function
“HoLd”
“non”
/
“PrEV”
Previous or non
non
Non hold
6.1.3
Burnout
function
“nAMU”
“non”
/
“Er
.Lo”
/
“Er
.HI”
Hi/Low/non
non
Non burnout
6.1.4
O
2
conc.
Hi/Low
alarm
“ALM”
“oFF”
Non alarm
oFF
Non alarm
6.1.5
“ALL”
“AL.Lo” “AL.HI”
Low limit Hi limit
“AL.HI”
Hi limit only
“AL.Lo”
Low limit only
Calibration gas
concentration
“SEt.C”
“1Pnt”
1 point
1000 ppm
Cal-gas conc.
(Note
2)
6.1.6
“2Pnt”
2
points
“Zero”
Zero-gas conc.
100.0 ppm
(Note
3)
“SP
An”
Span-gas conc.
1000 ppm
(Note
4)
“ALL”
3 points
“10”
10 ppm gas conc.
10.00 ppm
(Note
5)
“1000”
1000 ppm gas conc.
1000 ppm
(Note
6)
Output
smoothing
“SMoo”
“0
to
60”
Smoothing factor
.
0 sec
Non
(Note
7)
6.1.8
Multi-selector
“MLS”
“0”
/
“1”
/
“2”
/
“3”
Flow
path
No.1
to
3
“0”
Non use
6.1.9
(Note
1)
Auto
Range
between
0-10
ppm,
0-100
ppm,
0-1000
ppm,
0-1%,
0-10%,
and
0-100%.
(Note
2)
Calibration
gas
concentration
input
range:
0.90
ppm
to
12.00%,
35.00%
to
100.0%.
Gas
concentration
can
also
be
set
from
“CAL”
calibration.
(Note
3)
“ZEro”
input
range:
1.00
ppm
to
100.0%.
Gas
concentration
can
also
be
set
from
“CAL”
calibration,
(Note
4)
“SP
An”
input
range;
1.00
ppm
to
100.0%.
Input
is
not
possible
if
the
smallest
span
is
less
than
20%
of
the
range
after
inputting
“SP
An.”
Gas
concentration
can
also be set from “CAL” calibration,
(Note
5)
Calibration
gas
concentration
input
range:
8.00
to
20.00
ppm.
Gas
concentration
can
also
be
set
from
“CAL”
calibration,
(Note
6)
Calibration
gas
concentration
input
range:
800
to
2000
ppm.
Gas
concentration
can
also
be
set
from
“CAL”
calibration,
(Note
7)
If
the
output
smoothing
is
set,
the
smoothed
value
becomes
an
instantaneous
value
which
is
displayed
and
output
in
analog
format.
The
smoothed
value
is
also
used
as
the
instantaneous
value
(measured
value)
for
checking
concentration
alarms
and
RS232
communication.
(Note
8)
Set
parameters
will
not
be
erased
by
turning
the
power
off
and
on.