HART Communication
6
TI 12A01F01-61EN
Nov. 20, 2018-00
FLXA402 use the following extended Units code for each unit
Unit
Unit Code
rH
240
S/m
241
µS/m
242
S/cm
243
Ohm m
244
MOhm m
245
%SAT
247
1.2.4
Diagnostics Command (Command 48)
HART Status Group CMD48
byte
bit
HART Alarm Name
Alarm
Number NE107
Device Status
-
0 PV Out of Limit
-
S
1 Non-PV Out of Limit
-
S
2 Loop Current Saturated
-
S
3
Loop Current Fixed
-
N
4 More Status Available
-
N
5 Cold Start
-
N
6
Configuration Changed
-
N
7 Device Malfunction
-
N
Device Specific Status 0
0
0 Hardware failure
001
F
1 Internal com. error
002
F
2 IO mod. param. read error
003
F
3 Com. mod. param. read error
004
F
4 CPU param. read error
005
F
5 Reserved
006
N
6 Reserved
007
N
7 Reserved
008
N
Device Specific Status 1
1
0 VGB calc. error
009
F
1 Ratio calc. error
010
F
2 Passage calc. error
011
F
3 Rejection calc. error
012
F
4 Deviation calc. error
013
F
5 Script error
014
N
6 Reserved
015
N
7 Reserved
016
N
Device Specific Status 2
2
0 mA1 output burn out
017
F
1 mA2 output burn out
018
F
2 mA3 output burn out
019
F
3 mA4 output burn out
020
F
4 mA1 saturation
021
S
5 mA2 saturation
022
S
6 mA3 saturation
023
S
7 mA4 saturation
024
S
Device Specific Status 3
3
0
AI exceeds limit
025
S
1 AI out of range
026
F
2
Expiry time exceeded
027
M
3 Reserved
N
4 Wash response failure
029
M
5 Reserved
N
6 Converter Error Simulation
031
C
7 Fail safe occur
032
F