HART Communication
3
TI 12A01F01-61EN
Nov. 20, 2018-00
n
Installing DD File
For how to install the DD file, see the respective documentation for the setup tool you use.
For FieldMate R3.04.00+Device Files R3.09.00, installing DD File is not required.
n
Connecting Setup Tool
Connect HART modem at the both ends of the load resistance which is 250 ohms or larger and is
installed between the setup tool terminals and mA1.
You can connect the setup tool to any interconnection terminals such as those in the central
control room, a converter’s junction box, or somewhere within the transmission loop.
1.2
Functions Available via HART
Communication
FLXA402 via Hart communication enables the following settings: Error settings (4.7), mA output
settings (4.3), Contact input settings (4.2), HART setting (4.6.2). Information on the sensor
settings or calibration are available as Read only. The numbers in the parentheses correspond
the section number in the User’s Manual
Correspondences between menu structure and FLXA402 front panel display
+– Converter Menu
……… Shows the information of the Slot 1 sensor in FLXA402
+– HOLD
……… Enables Manual HOLD
+– WASH
……… Not operational
+– Detail
……… Read only
+– Trend
……… Disables reference
+– Other
……… Disables reference
+– Reset
……… Disables reference
+– Setting
……… Setting changeable: ma Output, ma Input, HART
+– Lang
……… Disables reference
+– Alarm
……… Disables reference
+– sensor Menu
……… Displays information of sensors on Slot1
+– Detail
……… Read only
+– Calibration
………
Not executable
+– Reset Wellness ………
Not executable
+– Reset
………
Not executable
+– Setting
……… Read only
Nov. 20, 2018-00