125
IM FG410-01EN
Trigger signal
Gate on
Gate off
Internal gate signal
Burst sync signal
Output signal
Oscillation
Stop
Stop
Gate on
Gate off
Oscillation
i) To execute oscillation in half-cycle units: Set oscillation stop unit to half cycle
To stop oscillation in half-cycle units, set [OscStop] (oscillation stop unit) on the 2nd page of the
setting screen to [HalfCycle] (half cycle). Normally [Cycle] (1 cycle) is set. When [Cycle] (1 cycle) is set,
oscillation is done in integer cycles.,
j) Usage of stop level
The level when oscillation is stopped is normally set with the oscillation start/stop phase, but it can
also be set independently from that with a proportion of the amplitude full scale.
Set [StpLvl] (stop level) on the 2nd page of the setting screen to [On] (on) and set the level with a %
value.
k)
Noise triggered gate oscillation
Noise having no cycle, the gate-on period is the oscillation period, and the gate-off period is the period
stop interval. Further, since noise has no phase, the stop level setting is always enabled.
4.9 Burst Setting and Manipulation
Содержание FG420
Страница 10: ...VIII IM FG410 01EN...
Страница 165: ...151 IM FG410 01EN External dimensions FG410 Unit mm 7 17 General Characteristics...
Страница 166: ...152 IM FG410 01EN 7 SPECIFICATIONS External dimensions FG420 Unit mm...
Страница 167: ...153 IM FG410 01EN RACK MOUNTING DIMENSIONS Inch rack mounting dimensions for 1 unit Unit mm 7 17 General Characteristics...
Страница 168: ...154 IM FG410 01EN 7 SPECIFICATIONS Inch rack mounting dimensions for 2 units Unit mm...
Страница 169: ...155 IM FG410 01EN Millimeter rack mounting dimensions for 1 unit Unit mm 7 17 General Characteristics...
Страница 170: ...156 IM FG410 01EN 7 SPECIFICATIONS Millimeter rack mounting dimensions for 2 units Unit mm...
Страница 171: ...157 IM FG410 01EN 7 17 General Characteristics...
Страница 173: ...159 IM FG410 01JA...
Страница 178: ......
Страница 179: ......