14
3.
PANELS AND I/O TERMINALS
IM FG410-01EN
3.2.3 External modulation/addition input (MOD/ADD IN)
When the modulation source is external, except for FSK and PSK, an external modulation
signal is input. In the case of the FSK and PSK, the external trigger input is used as
external modulation signal input.
When not used as external modulation signal input, MOD/ADD IN may be used as the
external addition signal input. The gain during external addition is either ×2 or ×10.
External modulation input:
External addition input:
■
Input Specifications
3.2.4 External trigger input (TRIG IN)
TRIG IN may be used as external trigger input in the following cases. The polarity setting
can be changed.
• Single-shot sweep start trigger
• Gated single-shot sweep start trigger
• Trigger burst oscillation start trigger
Input voltage
±1 V full scale
Maximum allowed input
±2 V
Input impedance
10 k
Ω
Input frequency
During modulation: DC to 25 kHz
During addition: DC to 10 MHz (
−
3 dB)
Signal GND
Same potential as same channel waveform output, insulated from
housing (42 Vpk max.).
FG420: Insulation also between channels (42 Vpk max.)
CH1
MOD/ADD
IN
[Insulated from
housing]
MOD/ADD
IN
[Insulated from housing]
CH2
MOD/ADD
IN
[Insulated from
housing]
FG420
FG410
CH2
TRIG
IN
TRIG
IN
CH1
TRIG
IN
FG420
FG410
Содержание FG400
Страница 1: ...User s Manual FG400 Arbitrary Function Generator User s Manual Basic IM FG410 01EN 1st Edition ...
Страница 8: ...VI IM FG410 01EN ...
Страница 159: ...7 17 General Characteristics 147 IM FG410 01EN External dimensions FG410 Unit mm Rear view ...
Страница 160: ...148 7 SPECIFICATIONS IM FG410 01EN External dimensions FG420 Unit mm Rear view ...
Страница 161: ...RACK MOUNTING DIMENSIONS 149 IM FG410 01EN Inch rack mounting dimensions for 1 unit Unit mm RACK MOUNTING DIMENSIONS ...
Страница 163: ...RACK MOUNTING DIMENSIONS 151 IM FG410 01EN Millimeter rack mounting dimensions for 1 unit Unit mm ...
Страница 169: ...157 IM FG410 01EN ...
Страница 171: ......