IM 12B06K07-E-E

Figure 4: Program sequence

Cleaning I/1

Cleaning with 1. cleaning solution, e.g water.
„Cleaning I“ valve and drain valve are opened 
and closed again after the preset time (10…300 
seconds).

By opening the „Cleaning I“ valve, a delivery pump 
can be triggered via an outlet trigger (no. 21 + 22). 

Cleaning II

Cleaning with 2. cleaning solution, e.g. acid.

„Cleaning II“ valve and drain valve are opened 
and closed again after the preset time (0…300 
seconds). If this time is set to "0", this program step 
is left out completely.

By opening the “Cleaning II” valve, a delivery pump 
can be triggered via an outlet trigger (no. 23 + 24). 

CLeaning II RT

Exposure time for 2. cleaning solution

Cleaning II RT is a reaction time for the 2. cleaning 
solution and provided so that the cleaning acid, 
which was put into the rinsing chamber, is able to 
react, for example.
  
All cleaning valves and the drain valve stay closed. 
After the preset time (0…300 seconds), the next 
program step follows. If this time is set to “0”, this 
program step is left out completely.

CLeaning I/2 

Cleaning with 1. cleaning solution, e.g. water.
“Cleaning I” valve and drain valve are opened and 
closed again after the preset time (0…300 seconds).

If this time is set to “0”, this program step is left out 
completely.

By opening the “Cleaning I” valve, a delivery pump 
can be triggered via an outlet trigger (no. 21 + 22).

Pause

If the sensor is not to be moved again into the 
process directly after cleaning, the pause time is 
activated.

This is particularly recommended if the measuring 
medium is especially aggressive and therefore the 
lifetime of the sensor is especially short. In this case, 
short measuring intervals and long pause times can 
minimize the dipping time of the sensor and thus 
increase the lifetime.

The sensor remains in the rinsing chamber, all 
cleaning valves and the drain valve remain closed. 
After the preset time (0…999 minutes), the sensor 
moves to the measurement position. 
If the time is set to “0”, this program step is left out 
completely.

Measure 

Measuring interval at parameterization “loop” or 
“loop + trigger”.

The sensor is moved to the measurement position 
and stays there for the preset time (1…999 
minutes). This time can be interrupted in case of 
parameterization “loop + trigger” by an external 
trigger. After this time has elapsed, the next 
cleaning cycle begins.

Measure

Clean I/1

Clean II

Clean II/RT

Clean I/2

Pause

Measure

Start

Loop / Event / Trigger

End

Trigger OFF

3.3 Cleaning program

As soon as a cleaning program is started (see 3.1), 
the following functions are run one after the other:

Содержание Exner RF20C

Страница 1: ...IM 12B06K07 E E 1st Edition Instruction Manual RF20C EXmatic 460 Control unit...

Страница 2: ...ery 13 4 1 Delivery scope 13 4 2 Delivery inspection 13 5 Assembly 15 5 1 Wall mounting 15 5 2 Electric connections 16 5 3 Pneumatic connections 17 5 4 Rinse valve connections 18 6 Parameterization 19...

Страница 3: ...he connected retractable fitting is suited and able to function Carefully read the operating instructions of the used retractable fitting and observe the notes on safety contained in it Take appropria...

Страница 4: ...is the cleaning times measuring intervals and the starting times can be parameterized and adjusted to the respective specification Input The control monitors the respective position indication of the...

Страница 5: ...can be started via an external trigger e g from the pH transmitter The control unit RF20C EXmatic 460 is a completely independent control unit and can principally be operated without any connection to...

Страница 6: ...Event In addition further cleaning processes can be started via an external trigger and the retractable fitting can be kept in the cleaning position This function is used if the sensor is to be water...

Страница 7: ...0 300 seconds If this time is set to 0 this program step is left out completely By opening the Cleaning I valve a delivery pump can be triggered via an outlet trigger no 21 22 Pause If the sensor is...

Страница 8: ...eep the control unit in the packaging There it is protected best until being assembled 4 2 Delivery inspection Before releasing the control unit for assembly you must ensure that packaging and unit ar...

Страница 9: ...only use approved tools Mount the control cabinet as shown below Figure 6 Wall mounting CAUTION DANGER WARNING 15 5 1 Wandmontage Stellen Sie sicher dass gen gend Arbeitsraum f r den Betrieb der Armat...

Страница 10: ...re 7 Connection diagram There is the risk for life and limb if the voltage supply is not disconnected Die Armaturensteuerung muss stromfrei geschaltet und die Druckluft drucklos sein Ist die Spannungs...

Страница 11: ...4 Output 24VDC max 80mA Clean Pump II 31 Output 24VDC max 100mA measurement 32 Output 24VDC max 100mA measurement 33 Output 24VDC max 100mA service Hold 34 Output 24VDC max 100mA service Hold 35 Outpu...

Страница 12: ...Manual Mode you reach the level for manual operation see chapter 6 3 When you select Parameters you reach the parameterization see chapter 6 4 When you select System you reach the system control see...

Страница 13: ...n of the function please refer to chapter 3 6 4 Parameters You can parameterize the times and the functions of the control in the parameters menu If you have chosen Parameter in the main menu or durin...

Страница 14: ...shows the currently set starting time You can change this value by pressing the soft keys Keep one soft key pressed for 2 seconds and then the value starts changing automatically By pressing the NEXT...

Страница 15: ...you switch from hours to minutes and seconds and back By confirming with Enter you save the changed value By pressing the ESC button you leave the menu without saving Set Date In the system mode Set...

Страница 16: ...display is shown Change Code Code Manual Mode Code Parameter Code System UP DOWN If you change or forget a pass code you block the system access Write the set values down and store them at a safe pla...

Страница 17: ...Yearly inspections Check the compressed air connections for tightness Check the multi hose connections for tightness and for a stable fit at the fitting and the control valves Retighten the clamps in...

Страница 18: ...external trigger 24V DC internal power supply for potential free contact Output for external relais Cleaning Pump I and Cleaning Pump II 24V DC 80mA max Output for status and alarm contacts 24V DC 100...

Страница 19: ...el cabinet 2 083 70 002 pipe mounting cabinet plastic stainless steel 2 083 70 003 Cleaning valves Accessory Order number Membrane valve 1 valve for cleaning solution or drainage 2 095 70 001 2 valves...

Страница 20: ...Netherlands www yokogawa com eu YOKOGAWA ELECTRIC CORPORATION World Headquarters 9 32 Nakacho 2 chome Musashino shi Tokyo 180 8750 Japan www yokogawa com YOKOGAWA CORPORATION OF AMERICA 2 Dart Road Ne...

Отзывы: