YOKOGAWA EXA100 Series Скачать руководство пользователя страница 2

IM 12D11A01-11E

1st Edition: Apr. 2003

4. Measurement

Sample Temperature Conditions

SC10XB Conductivity Sensor : 70 

8

C or lower

Note: Specification requirements vary depending on sample composition.

• The SC10XB can be used in conjunction with a dedicated adapter.

5. Maintenance

• Cleaning the Electrode

  

<Periodical cleaning, sample dependent>

• Cleaning the Electrode

6. Error Messages, Error checking flowchart

• An error message during calibration

For details on error messages, refer to Chapter 8 of the SC100 Panel Mount Converter's User's Manual, IM 12D11A01-01E

Clean the electrode.

Check connections.

NO

NO

NO

NO

NO

The electrode is wet.

The electrode is dirty.

YES

YES

YES

YES

YES

The electrode has 

deteriorated. Replace 

the sensor with a new 

one.

Water

Electrode

Cotton wab or Soft Material

Check connections.

Clean the electrode.

NO

NO

NO

The temperature 

sensor is disconnected 

or short-circuited.

The electrode is dirty.

Cable connections 

are not appropriate.

YES

YES

YES

The electrode has 

deteriorated. 

Replace the sensor 

with a new one.

Zero Calibration (Air Calibration: Eliminating an error resulting from wiring capacity)

Refer to Section 6.4 of the SC100 Panel Mount Converter's User's Manual, IM 12D11A01-01E.

Replace the 

sensor.

 

• An error message during measurement

NO

YES

Dry it thoroughly.

Water

Calibration 
Display Top

Replace the sensor 
in the process.

Calibration result 
is reflected to the 
measured value.

Main display stops 
flashing and 
sub-display is flashing.

Press [ENT] key once.

Press [

n

] key 

for 3 sec.

Measurement 
Display

Wait until air 
calibration has 
been complete 
(60 sec).

"Air" in sub-display 
is flashing.

Press [

,

] key 

for 3 sec.

Measurement 
Display

Press [ENT]
 key once.

Rinse the conductivity 
sensor well with water 
or appropriate solution.

Press [

n

] key.

µS/cm

mS/cm

EXA

SC100

S1

S2

S3

S4

mS/cm

µS/cm

EXA

SC100

ENT

S1

S2

S3

S4

SET

CAL

MEAS

CANCEL

µS/cm

mS/cm

CANCEL

MEAS

CAL

SET

S4

S3

S2

S1

ENT

SC100

EXA

µS/cm

mS/cm

S4

S3

S2

S1

SC100

EXA

Press [ENT
key once.

In the air

µS/cm

mS/cm

EXA

SC100

S1

S2

S3

S4

µS/cm

mS/cm

S4

S3

S2

S1

SC100

EXA

µS/cm

mS/cm

EXA

SC100

S1

S2

S3

S4

Data updated

µS/cm

mS/cm

EXA

SC100

S1

S2

S3

S4

Main display is 

flashing and "WAI" in 

sub-display is flashing.

µS/cm

mS/cm

EXA

SC100

S1

S2

S3

S4

"Str" in sub-display 
is flashing.

"CAL" in sub-display 
is flashing.

Press [ENT
key once.

Error message C-2, C-3, or C-4 appears 
during calibration.

Error message E-1, E-2, E-3, or 
E-5 appears during measurement.

Use standard solution with 
correct concentration.

Cable connections 

are not appropriate.

Dry the electrode in zero (air) 
calibration.

When error message 
C-1 appears, bring the 
standard solution within 
the operating temperature.

* Correct temperature 
   coefficient should be set when error message C-4 appears.

If an E-4 error message 
appears, the converter has malfunctioned.
Contact your sales representative or the nearest service office.

Converter failure has 

been detected.

Replace the 

converter.

Converter failure 

has been detected.

Replace the 

converter.

The concentration 

of standard solution is 

incorrect.