IM 12D7B3-E-E
Parameter setting 5-9
Code Display
Function
Function detail
X
Y
Z
Default values
Parameter specific functions
01
*SC.RES
Select main parameter Conductivity
0
0
Cond.
Resistivity
1
02
*4-ELEC
Select 2/4-EL system 2-Electrode measurement system
0
0
2-El.
4-Electrode measurement system
1
03
*0.10xC
Set cell constant
Press NO to step through choice of
0.100 cm
-1
multiplying factors on the second display.
0.10xC
1.00xC
0.10xC
10.0xC
100.xC
0.01xC
Press YES to select a factor
Use >, ^, ENT keys to adjust MAIN digits
1.000
04
*AIR
Zero calibration
Zero calibration with dry cell connected
*START
Press YES to confirm selection
*”WAIT”
Press YES to start, after briefly displaying
*END
“WAIT”, *END will be displayed
Press YES to return to commissioning
mode
05
*POL.CK
Polarization check
Polarization check off
0
1
On
Polarization check on
1
06-09
Not used