IM 01C25H01-01E
8-2
8. GENERAL SPECIFICATIONS
NORMAL OPERATING CONDITION
(Optional features or approval codes may affect
limits.)
Ambient Temperature Limits
–40 to 60
°
C (–40 to 140
°
F)
–30 to 60
°
C (–22 to 140
°
F) with LCD display
(Note : The ambient temperature limits must be within the
fill fluid operating temperature range, see table 8.1.)
Process Temperature Limits
See table 8.1.
Ambient Humidity Limits
0 to 100% RH
Working Pressure Limits
See table 8.1.
For atmospheric pressure or below, see figure 8.1 and 8.2.
2.7 (0.38)
100 (14.5)
0.1(0.014)
1(0.14)
10 (1.4)
-50
0
50
100 150 200 250 300 350
F0801.EPS
Process temperature
for fill fluid code B
Process temperature
for fill fluid code A
Process temperature
for fill fluid code C
Flange max.
working
pressure
Atmospheric
pressure
Transmitter ambient
temperature range
(for fill fluid code
A, B)
Process Temperature
(
C)
Working
pressure
kPa abs
(psi abs)
Figure 8.1
Working Pressure and Process Temperature
(Fill fluid : silicone oil for general and high
temperature use)