![YOKOGAWA DpharpEJX vigilantplant EJX Series Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/yokogawa/dpharpejx-vigilantplant-ejx-series/dpharpejx-vigilantplant-ejx-series_manual_3728307026.webp)
IM 01C25T01-01E
3-18
3. OPERATION
(3) Sensor Trim for Static Pressure
For the EJX differential transmitters, zeroing and full
sensor trim of the static pressure is performed in the
same way as with the primary process variable (PV).
Note that the static pressure sensor trim should be done
only after trimming the PV.
(4) Reset Trim Adjustment to Factory Setting
The
Clear P snsr trim
and
Clear SP snsr trim
commands can reset the trim adjustment to the initial
calibrated values that were set. The amount of the
adjustment by the external zero-adjustment screw is
returned to the initial setting as well.
3.5.8
Trim Analog Output
Fine current output adjustment is carried out with
D/A
trim
or
Scaled D/A trim
.
•
D/A Trim
D/A trim
is to be carried out if the calibration digital
ammeter does not exactly read 4.000 mA and 20.000
mA with an output signal of 0% and 100%.
•
Scaled D/A Trim
Scaled D/A trim
is to be carried out if the output is
adjusted using a voltmeter or a meter whose scale is
0 to 100%.
1
F0330.EPS
Select the
D/A trim
item.
EJX:YOKOGAWA
Analog output trim
1 D/A trim
2 Scaled D/A trim
3 Clear D/A trim
HELP
SAVE
HOME
OK
EJX:YOKOGAWA
WARN-Loop should be
removed from
automatic control
DEL
SET
ABORT
OK
2
Press
OK (F4)
.
EJX:YOKOGAWA
Connect reference
meter
HELP
DEL
ABORT
OK
3
Connect the ammeter (
±
1
µ
A is
measurable), and press
OK (F4)
.
EJX:YOKOGAWA
Setting fld dev
output to 4mA
HELP
ABORT
4
Press
OK (F4)
, and the transmitter
outputs a 0% output signal.
DEL
OK
(OK)
(OK)
(OK)
EJX:YOKOGAWA
Enter meter value
4.000
4.115
HELP
ESC
5
Enter the read value
4.115
of the
ammeter, and press
ENTER (F4)
.
(The output of the transmitter
changes.)
DEL
ENTER
(ENTER)
‘4 . 1 1 5’
Example 1: For an adjustment using an ammeter (
±
1
µ
A is
measurable)
1. Device setup
2 Diag/Service
3. Calibration
2. Analog output trim
1. D/A trim
Ammeter reading: 4.115