IM 01C21F01-01E
8-9
8. BRAIN TERMINAL BT200 OPERATION
(4) Output Signal Low Cut Mode Setup
(D10: LOW CUT, D11: LOW CUT MODE)
Low cut mode can be used to stabilize the output
signal near the zero point. The low cut point can be
set in a range from 0 to 20% of output. (Hysteresis of
cut point:
±
1%)
(%)
50
10
10
0
50
Input(%)
Output
LOW CUT at 10%
F0816.EPS
• Example: Change the low cut setting range
from 5% to 10%, and the low
cut mode from LINEAR to ZERO.
FEED
NO
OK
SET
D11:LOW CUT MODE
ZERO
Use the or key
to select ZERO.
Press the key twice to
enter the setting.
Press the (OK) key.
F0817.EPS
FEED
NO
OK
SET
D10:LOW CUT
10.0 %
ESC
SET
D11:LOW CUT MODE
LINEAR
< LINEAR >
< ZERO >
DATA
DIAG
PRNT
ESC
PARAM
D10:LOW CUT
10.0 %
D11:LOW CUT MODE
ZERO
D20:DISP SELECT
NORMAL %
CLR
ESC
SET
D10:LOW CUT
5.0 %
+ 10.0
Set “10.”
Press the key twice to
enter the setting.
Press the (OK) key.
Next, the [D11: LOW CUT MODE]
setting panel is displayed.