3. INSTALLATION OF DO71

 

Using the sensor adapters the DO71 

optical dissolved oxygen sensor can be 

assembled into the complete Yokogawa 

flow and immersion fitting program: 

Flow fitting FF20, FF40, PB350G/PB30 an

-

gled floating ball holder, PB360G vertical 

floating ball holder, or DOX8HS submer

-

sion type holder. Flow and immersion 

fitting program allows sensor to be 

immersed or submerged to the optimum 

point to obtain precise measurements. 

Section 3.7 of this manual describes 

adapters for mounting sensor directly in 

process or existing fitting program.

3.1 Installation in sensor holder

An adaptor (optional) is necessary to as-

semble DO71 into the DOX8HS submer-

sion type holder (Fig. 2), PB350G angled 

floating ball holder (Fig. 3) or PB360G 

vertical floating ball holder (Fig. 4).

3.2 Preparing the sensor for use 

Make sure that the Oxygen Exchange 

Cap (OEC) is screwed tightly to the probe 

body and did not work loose while han-

dling the probe.

3.3 Mounting the sensor 

Choose a holder / fitting that suits the 

liquid to be measured and the measure-

ment location. For detailed sensor instal-

lation procedure to sensor holder, please 

refer to section 3.5 and 3.6.

Note:

 Avoid choosing a location where 

the solution is unevenly distributed, 

since heavy hunting may result in the 

measured value. Also, avoid locations 

where bubbles occur frequently.

Note that, if a measured solution run-

ning at a high flow rate contains particles 

of sand, the sensor membrane may be 

damaged.

Fig. 2:

 DOX8HS submersion type holder

Fig. 4: 

PB360G vertical floating ball 

holder

Fig. 3: 

PB350G/PB30 angled floating ball 

holder

10

IM12J06J01-00EN-P

Содержание DO71

Страница 1: ...Instruction Manual Model DO71 Optical Dissolved Oxygen Sensor IM12J06J01 00EN P 3rd edition...

Страница 2: ...ereafter simply referred to as YOKOGAWA when replacing parts or consumables Modification of the product is strictly prohibited The following safety symbols are used on the product as well as in this m...

Страница 3: ...se but rather to describe the functional details of the product No part of the Instruction manuals may be transferred or reproduced with out prior written consent from YOKOGAWA YOKOGAWA reserves the r...

Страница 4: ...Installation in sensor holder 10 3 2 Preparing the sensor for use 10 3 3 Mounting the sensor 10 3 5 Installation procedure for the sensor to PB350G PB360G 11 3 6 Installation procedure for the sensor...

Страница 5: ...a products and parts are guar anteed free from defects in workmanship and material under normal use and ser vice for a period of typically 12 months from the date of shipment from the man ufacturer In...

Страница 6: ...uction Year code Year Year code Year Year code 2014 P 2026 3 2015 R 2027 4 2016 S 2028 5 2017 T 2029 6 2018 U 2030 7 2019 V 2031 8 2020 W 2032 9 2021 X 2033 A 2022 Y 2034 B 2023 Z 2035 C 2024 1 2036 D...

Страница 7: ...lbs 0 095 lbs 0 015 lbs feet Sensor mounting system Pressfit with optional adapter Cable Permanent shielded cable stripped wire end 5 mm for terminal connection 2 4 Performance Measuring range 0 to 22...

Страница 8: ...97 0 65 g lbs DO71 10m 32 80 ft 232 0 51 g lbs 391 0 86 g lbs DO71 20m 65 61 ft 420 0 92 g lbs 615 1 35 g lbs OEC CA CB CC 6 0 01 g lbs 147 0 32 g lbs 2 6 Environmental conditions Storage temperature...

Страница 9: ...010 1 2010 A1 2019 NEN EN IEC 61010 2 030 2010 C1 2011 RoHS Directive 2015 863 EU Commission Delegated Directive EU 2015 863 amending Annex II as regards the list of restricted substances and by apply...

Страница 10: ...oating ball holder Fig 3 or PB360G vertical floating ball holder Fig 4 3 2 Preparing the sensor for use Make sure that the Oxygen Exchange Cap OEC is screwed tightly to the probe body and did not work...

Страница 11: ...rough the cable and put O RING on GLAND to touch the flange of ADAPTER The installation is correct if the distance between the head of the sensor and the end face of ADAPT ER is less than 2mm By posit...

Страница 12: ...and the end face of ADAPTER is less than 87mm By positioning the sensor the measure ment point can be the same even after the reinstallation with maintenance works Also the installation position gets...

Страница 13: ...01 00EN P Figure 11 DO71 installation in Yokogawa FF20 fitting6 using the adapter K1531JA sensor adapter set and adapter to fitting K1520JN K1500DV or K1520JP Adapter set K1531JA Adapter K1520JN K1500...

Страница 14: ...allation in Yokogawa FF40 fitting6 using the adapter K1531JA sensor adapter set to fit in PG13 5 fitting adapter K1523JA Noryl K1523JC SS Adapter set K1531JA PG13 5 adapter K1523JA Noryl K1523JC SS 14...

Страница 15: ...le lengths as described in table 3 Sensor and O ring dimensions are described in Fig 13 Table 3 Sensor model code and cable length IM12J06J01 00EN P 15 Model code L mm L ft DO71 W 106 AA 03 N Y 3000 1...

Страница 16: ...DO71 sensor cable connection to FLXA402 refer to the Instruc tion manual for FLXA402 4 Wire Converter IM12A01F01 02EN Fig 14 D6 module terminal assignment for DO71 dissolved oxygen sensor Wire Pin no...

Страница 17: ...nect the DO71 to the protective ground at the installation site Assign the device housing and cable shield to protective ground Especially the length of the cable can influence the ground connection q...

Страница 18: ...ately one minute to dissolve Na2SO3 and to ensure that the water is oxygen free To prepare oxy gen free water you also can use sodium dithionite Na2S2O4 Fill the calibration solution cal 0 in the vess...

Страница 19: ...IM12J06J01 00EN P 19 Fig 17 Example of Final inspection protocol...

Страница 20: ...he dirt Wash the oxygen exchange cap with clean water Before every calibration check that the oxygen exchange cap is not loose If it is loose retighten it Cleaning procedure 2 H2O2 acidic agents HCL H...

Страница 21: ...DO71 Optical Dissolved Oxygen sensor Range W Wide range 0 22 5 mg L Insert length 106 106 mm Type AA General purpose Connection type 3 Permanent cable 3 meters 5 Permanent cable 5 meters 10 Permanent...

Страница 22: ...6 PCS K1500GR O RINGS SILICON 10 77x2 62 8PCS K1524AA O RING SIL 10 77x2 62 SLIDE RING 1PC K9679WJ Gasket lifetime parts for K9679WA WB and WC K1531JA Adapters Adapter for mounting in Yokogawa fittin...

Страница 23: ...on Oxygen Exchange Cap membrane Compatibility Aqueous solutions ethanol methanol No cross sensitivity to pH 1 14 CO2 H2S SO2 ionic species Cross sensitivity to Chlorine gas Organic solvents such as p...

Страница 24: ...Subject to change without notice Copyright Printed in The Netherlands 03 2211 IM12J06J01 00EN P...

Отзывы: