<2. Installing the Sensor and Connecting the Cable>>
2-1
IM 12J05D04-01E
2. Installing the Sensor and Connecting
the Cable
2.1
Preparing for Installation
Protection caps are mounted at both ends of the D70G to prevent the sensor cap or sensor plugs
from being damaged. When installing the sensor, remove these protection caps.
n
Attaching/Detaching the sensor cable
Attachment
●
Rotating the sensor connector so that its recessed portions will agree with the salient portions
inside the cable connector, insert the sensor connector into the cable connector.
Note: A weak pressure is sufficient for inserting it. If a strong pressure is needed to insert it, the connection position
will be incorrect. Check the position and re-try.
● After inserting the sensor connector completely, grip the portion “A”, and rotate the cap nut
(“B” portion) clockwise until the nut cannot rotate.
Detachment
●
Gripping the portion “A”, and rotate the cap nut (“B” portion) un-clockwise.
● After loosing the nut completely, grip and pull the plastic portion of the sensor from the cable
connector.
A
B
A
B
Recess for positioning
the sensor and cable
(Sensor portion)
(Sensor portion)