IM 12J05C01-01E
1-2 Appendix
A-1-2 Configuration Checklist for DO202
Standard configuration
Options
Reference for change
Measured variable(s)
Primary inputs
DO and Temp
DO units
ppm
ppb. % saturation
code 56
Temperature unit
Celsius
Fahrenheit
code 11
mA Outputs
Analog output
4-20mA for ppm
ppb or %sat
code 56, 31
Output linearization
disabled
ppm, ppb or %sat
code 35
Burn out
disabled
burn low (3.6) / high (21) on mA output code 32
Communication
Digital interface
HART
disabled
code 60
Communication software external
HHC
contact factory
Interface
Variables on display
ppm and temp
%sat, °C, SE, ZR, REL
“display”
Password protection
disabled
for mains/ comm./ serv level
code 52
Autoreturn
return to measure in 10 minutes
enable or disable
code 50
Add. functions in MAINT disabled
manual pressure
code 51
Diagnostics
Check on Slope
active
1-1999 na/ppm
code 22
Check on zero
disabled
enable or disable
code 02, 21, 22
Check on stability
0.1 ppm per 60 s
0-50 ppm in 5-600 s
code 20
Compatibility
DO sensor
SM31
DO30G
code 01, 10
Temperature sensor
22k NTC
Pt 1000
code 10
Sensor principle
polarographic
galvanic
code 01, “wiring”
Manual temp. comp.
disabled
disable or enable
code 13
Special Features
Salinity compensation
disabled
comp. for 1-99.9 ppt NaCI
code 04
Temperature calibration none
adjustment ±7.5 °C or ±13.5°F
code 12
Zero calibration
disabled
disable or enable
code 21
Atm. press. compensation manual
Soft fail alarm
disabled
possible for E1..E2, E7..E9, E16
code 53
Logbook
disabled
2 volumes of 50 events
code 61,62
mA
mA
Through HART the user has access to the process
variables:
PV: User selected process value: ppm, ppb or %
saturation
SV: Temperature
TV: % Saturation
QV: sensor current
Each of these process variables can be decoded into
current output or contact output through the HIM: Hart
Interface Module. Yokogawa offers standard a HIM
with two 0- 20 or 4- 20 mA outputs and three contact
outputs with 24 VDC power supply.
Full communication with the DO202 can be done
through various brands of HART Handheld termi-
nals (HHT) or through Pactware. This program and
the Yokogawa DTM library can be downloaded
from YOKOGAWA WEB sites free of charge by
Yokogawa customers.
Yokogawa recommends the use of MH-02 modem
for HART-RS232 conversion.
A-1-3 HART process variables
mA
Содержание DO202G
Страница 7: ...IM 12J05C01 01E LT LV PL EST SLO H BG RO M CZ SK...
Страница 11: ...IM 12J05C01 01E 1 4 Introduction...
Страница 31: ...IM 12J05C01 01E General Specifications 2 20...
Страница 37: ...IM 12J05C01 01E...
Страница 61: ...IM 12J05C01 01E 9 1 Spare Parts 9 SPARE PARTS See Customer Maintenance Parts List Figure 9 1 Exploded view...
Страница 65: ...IM 12J05C01 01E...
Страница 69: ...IM 12J05C01 01E 2 4 Appendix...
Страница 74: ...IM 12J05C01 01E Appendix 2 9...
Страница 78: ...IM 12J05C01 01E Appendix 2 13...
Страница 82: ...IM 12J05C01 01E Appendix 2 17...