5-125
IM DLM4038-17EN
Commands
1
2
3
4
5
6
App
Index
:LOGic:PODA:BIT<x>:TYPE
Function
Selects the threshold level of a bit of logic input
port A.
Syntax
:LOGic:PODA:BIT<x>:TYPE {CMOS1|
CMOS2|CMOS3|CMOS5|ECL}
:LOGic:PODA:BIT<x>:TYPE?
<x> = 0 to 7
Example
:LOGIC:PODA:BIT1:TYPE CMOS1
:LOGIC:PODA:BIT1:TYPE?
-> :LOGIC:PODA:BIT1:TYPE CMOS1
Description An error will occur if 16 bit input (/L16 option) is
not supported.
:LOGic:PODA:DESKew
Function
Sets or queries the deskewing of logic input port A.
Syntax
:LOGic:PODA:DESKew {<Time>}
:LOGic:PODA:DESKew?
<Time> = −100 to 100 ns (in 10-ps steps)
Example
:LOGIC:PODA:DESKEW −100
:LOGIC:PODA:DESKEW?
-> :LOGIC:PODA:DESKEW −100
Description An error will occur if 16 bit input (/L16 option) is
not supported.
:LOGic:PODA:HYSTeresis
Function
Sets or queries the hysteresis of logic input port A.
Syntax
:LOGic:PODA:HYSTeresis {HIGH|LOW}
:LOGic:PODA:HYSTeresis?
Example
:LOGIC:PODA:HYSTERESIS HIGH
:LOGIC:PODA:HYSTERESIS?
-> :LOGIC:PODA:HYSTERESIS HIGH
:LOGic:PODA_PODB?
Function
Queries all settings of logic input ports A and B.
Syntax
:LOGic:PODA_PODB?
Example
:LOGIC:PODA_PODB?
Description An error will occur if 16 bit input (/L16 option) is
not supported.
:LOGic:PODA_PODB:BITOrder
Function
Sets or queries the location of logic input ports A
and B.
Syntax
:LOGic:PODA_PODB:BITOrder {<String>}
:LOGic:PODA_PODB:BITOrder?
<String> = Up to 40 characters
Example
:LOGIC:PODA_PODB:BITORDER "ABC"
:LOGIC:PODA_PODB:BITORDER?
-> :LOGIC:PODA_PODB:BITORDER "ABC"
Description An error will occur if 16 bit input (/L16 option) is
not supported.
:LOGic:PODA_PODB:BUS2?
Function
Queries all bus settings of logic input ports A and
B.
Syntax
:LOGic:PODA_PODB:BUS2?
Example
:LOGIC:PODA_PODB:BUS2?
Description An error will occur if 16 bit input (/L16 option) is
not supported.
:LOGic:PODA_PODB:BUS2:ASSignment
Function
Sets or queries the assignment of a bus of logic
input ports A and B.
Syntax
:LOGic:PODA_PODB:BUS2:
ASSignment {<String>}
:LOGic:PODA_PODB:BUS2:ASSignment?
<String> = Up to 16 characters
Example
:LOGIC:PODA_
PODB:BUS2:ASSIGNMENT "ABC"
:LOGIC:PODA_PODB:BUS2:ASSIGNMENT?
-> :LOGIC:PODA_
PODB:BUS2:ASSIGNMENT "ABC"
Description An error will occur if 16 bit input (/L16 option) is
not supported.
:LOGic:PODA_PODB:BUS2:DISPlay
Function
Sets or queries the display on/off status of a bus
of logic input ports A and B.
Syntax
:LOGic:PODA_
PODB:BUS2:DISPlay {<Boolean>}
:LOGic:PODA_PODB:BUS2:DISPlay?
Example
:LOGIC:PODA_PODB:BUS2:DISPLAY ON
:LOGIC:PODA_PODB:BUS2:DISPLAY?
-> :LOGIC:PODA_PODB:BUS2:DISPLAY 1
Description An error will occur if 16 bit input (/L16 option) is
not supported.
:LOGic:PODA_PODB:BUS2:FORMat
Function
Sets or queries the display format (bus display) of
a bus of logic input ports A and B.
Syntax
:LOGic:PODA_
PODB:BUS2:FORMat {HEX|BINary}
:LOGic:PODA_PODB:BUS2:FORMat?
Example
:LOGIC:PODA_PODB:BUS2:FORMAT HEX
:LOGIC:PODA_PODB:BUS2:FORMAT?
-> :LOGIC:PODA_PODB:BUS2:FORMAT HEX
Description An error will occur if 16 bit input (/L16 option) is
not supported.