4-51
IM DLM3054-01EN
If the comparison condition is a ≤ Data ≤ b or “Data < a or b < Data,” the two reference values are
automatically adjusted so that the lower limit is less than or equal to the upper limit.
Byte Order (Endian), Sign (Sign), and Comparison Range (MSB/LSB)
This item is the same as that of the CAN bus trigger.
ID OR
The instrument triggers on the AND condition of SOF and multiple IDs.
The ID trigger condition is met when the ID matches any of the patterns that you set in ID1 to ID4.
• Trigger Condition (Condition Setup)
On the Condition Setup screen, set the PTYPE, ID, and other trigger conditions.
SOF
The instrument triggers on the start of a CXPI bus signal frame when the trigger condition is set to SOF only.
SOF of ID OR is always selected as a trigger condition.
ID
• with PTYPE
The instrument triggers on whether a frame contains PTYPE.
Yes: Triggers on the ID of a frame containing PTYPE
No: Triggers on the ID of a frame not containing PTYPE
X: Triggers on the ID regardless of the presence of PTYPE
• ID1 to ID4
You can set up to four IDs. IDs whose check boxes are selected are used as trigger conditions.
The ID trigger condition is met when the input signal ID bit pattern matches any of the specified ID bit
patterns. The method for setting the bit pattern of each ID is the same as with the ID/Data setting.
Wakeup/Sleep
The instrument triggers on any of the following: wakeup pulses, wakeup states, sleep frames, or sleep states.
• Trigger Type (Type)
Wakeup Pulse
When a pulse in the dominant period between 250 μs and 2500 μs is detected
Wakeup
When a transition from a state in which there is no clock after a sleep detection to a state
in which there is a clock is detected
Sleep Frame
When the sleep frame ID value is 1F (hexadecimal)
Sleep
When 5 ms elapses after a transition is made from a state in which there is a clock to a
state in which there is no clock
Source (Source)
Select the trigger source. After selecting the trigger trigger source, configure the bit rate, T Sample, clock
tolerance, HF rejection, trigger level, hysteresis, and other settings.
Source (Source)
Set the trigger source to one of the settings below.
CH1 to CH4
4 Triggering