IM DLM3054-01EN
Reference Range (Upper Level/Lower Level)
Set the upper and lower levels of the reference range.
Time Condition (Time Qualification)
Set whether the trigger source must be inside (IN) the specified range or outside (OUT) the specified range
(Window) and what kind of relationship must be established with the specified reference times (Time or Time1
and Time2) for the instrument to trigger.
None
Without a time condition
More than
When the time that the waveform stays inside or outside the
window is longer than the specified reference time (Time)
Less than
When the time that the waveform stays inside or outside the
window is shorter than the specified reference time (Time)
Inside
When the time that the waveform stays inside or outside the
window is longer than reference time Time1 and shorter than
reference time Time2.
Outside
When the time that the waveform stays inside or outside the
window is shorter than reference time Time1 or longer than
reference time Time2.
Timeout
When the time that the waveform stays inside or outside the
window is longer than the specified reference time (Time)
Reference Times (Time or Time1 and Time2)
The selectable range for Time or Time1 and Time2 varies depending on the time condition.
More than
Time: 2 ns to 10 s
Less than
Time: 4 ns to 10 s
Inside
Time1: 2 ns to [10 s - 2 ns], Time2: 4 ns to 10 s
Outside
Time1: 2 ns to [10 s - 2 ns], Time2: 4 ns to 10 s
Timeout
Time1: 4 ns to 10 s
Resolution: 2 ns
HF Rejection (HF Rejection), Noise Rejection (Noise Rejection)
Set these items for the trigger source.
These items are the same as those of the edge trigger.
Trigger Position (POSITION), Trigger Delay (DELAY), Trigger Hold-Off (Holdoff)
These items are the same as those of the edge trigger.
• The instrument may not trigger properly if the spacing between signals or the time detection width is less
than 1 ns.
• Under standard operating conditions, the time accuracy immediately after calibration is ±(0.002% of setting
+ 450 ps).