3-4
IM 701310-17E
3.4 GP-IB Interface Specifications
GP-IB Interface Specifications
Electrical and mechanical specifications: Conforms to IEEE St’d 488-1978
Functional specifications:
See table below.
Protocol:
Conforms to IEEE St’d 488.2-1992
Code used:
ISO (ASCII) code
Mode:
Addressable mode
Address setting:
The address can be set in the range from 0
to 30 on the GP-IB setting screen that is
played using the MISC menu.
Clear remote mode:
Remote mode can be cleared by pressing
CLEAR
except when the DL9000 has been
set to Local Lockout mode by the controller.
Functional specifications
Function
Subset Name Description
Source handshaking
SH1
Full source handshaking capability
Acceptor handshaking
AH1
Full acceptor handshaking capability
Talker
T6
Basic talker capability, serial polling, untalk on MLA
(My Listen Address), and no talk-only capability
Listener
L4
Basic listener capability, unlisten on MTA (My Talk
Address), and no listen-only capability.
Service request
SR1
Full service request capability
Remote local
RL1
Full remote/local capability
Parallel polling
PP0
No parallel polling capability
Device clear
DC1
Full device clear capability
Device trigger
DT0
No device trigger capability
Controller
C0
No controller capability
Electrical characteristics E1
Open collector
Data Rate
The response time when outputting waveform data is shown below as a reference.
Model:
701313
Controller:
PC (Pentium4 3.4 GHz, USB 2.0), OS (Windows XP Professional SP1)
Language used: Visual C++
Number of Data Points
Word Data
ASCII Data
2s500
Approx. 16 ms
Approx. 0.390 s
125000
Approx. 344 ms
Approx. 19.453 s
1250000
Approx. 3172 ms
Approx. 194.516 s
2500000
Approx. 6282 ms
Approx. 389.047 s
6250000
Approx. 15641 ms
Approx. 971.985 s