7-18
IM 701310-01E
Explanation
When the DL9000 detects the specified GO/NO-GO conditions, it activates an action-on-
trigger.
To set GO/NO-GO conditions, see sections 7.10 to 7.16.
Modes
You can select one of the following modes.
• Zone/Param
If you select Zone/Param, you must further specify the GO/NO-GO conditions by
selecting one of the following determination modes. For details, see section 7.10 to
7.15.
Wave
The DL9000 returns GO/NO-GO results based on a waveform zone. For
information about configuring this mode, see section 7.10.
Rect
The DL9000 returns GO/NO-GO results based on a rectangular zone. For
information about configuring this mode, see section 7.11.
Polygon
The DL9000 returns GO/NO-GO results based on a polygonal zone. For
information about configuring this mode, see section 7.12.
Parameter
The DL9000 returns GO/NO-GO results based on waveform parameters, XY
waveform parameters, or FFT parameters. To set waveform parameters, see
section 7.13. To set FFT parameters, see section 7.14. To set XY waveform
parameters, see section 7.15.
• Telecom Test
The DL9000 returns GO/NO-GO results based on a Telecom Test. For information
about configuring this mode, see section 7.16.
Setting the Logic to Apply to the GO/NO-GO Conditions
You can set up to four GO/NO-GO conditions. You can set each condition to one of the
modes listed above, and you can set the logic that you want to apply to the conditions
that you set.
AND The DL9000 returns a no-go result when all of the conditions are met.
OR
The DL9000 returns a no-go result when any of the conditions is met.
Test Count
Signal acquisition stops when the specified ACQ Count or Nogo Count is reached.
• ACQ Count
Set the signal acquisition count.
1 to 1000000
After the set number of signals has been acquired, acquisition stops.
Infinite (0)
Acquisition continues until you press the Abort soft key or START/STOP.
• Nogo Count
The DL9000 stops acquisition after it returns the set number of no-go results.
1 to 1000
Acquisition stops after the set number of no-go results are returned.
Infinite (0)
Acquisition continues until you press the Abort soft key or START/STOP.
Actions
The following four actions can be triggered in response to GO/NO-GO results.
• Beep
The DL9000 generates an alarm.
• Print/Save Display Data
The DL9000 prints or saves the display image data with or to the printer or storage
medium set by the Copy to option in the PRINT menu. (Set the Copy to option in the
PRINT menu to Printer to print to the internal printer. Set it to USB to print to a USB
printer).
• Save Measured Data
The DL9000 saves the measured data to the save destination specified in the FILE
menu. You can choose to save the data in binary, ASCII, or floating point format. The
data format is determined by the data type option in the FILE menu (see section 13.5).
7.9 Activating the Action-On-Trigger Function Using GO/NO-GO Results