2-38
IM DL850E-01EN
Pull Up (Pull Up)
You can choose whether to enable the pull-up function only when Preset is set to Pull-up 5V. You cannot
configure the pull-up setting when another preset is selected.
If you enable the pull-up function, set the input voltage to a value from 0 V to 5 V. If you apply a voltage that is
outside of this range, the internal protection circuit will cut off the pull-up circuit.
Internal Equivalent Circuit for When the Pull-up Function Is Enabled
+5 V
4.7
kΩ
990 kΩ
10 kΩ
Zoom Method (V Scale), Zooming by Setting a Magnification (V Zoom), Zooming by
Setting Upper and Lower Display Limits (Upper/Lower)
Offset (Offset)
Trace Settings (Trace Setup)
Linear Scaling (Linear Scale)
2 Vertical Axis